Efficacy of Somnodent® Appliance Therapy in Moderate to Severe Obstructive Sleep Apnea Patients Somnodent® OAs for Moderate to Severe OSAHS

Research Article

Efficacy of Somnodent® Appliance Therapy in Moderate to Severe Obstructive Sleep Apnea Patients Somnodent® OAs for Moderate to Severe OSAHS

  • Guillaume Buiret 1*
  • Frederic Chidiac 1

*Corresponding Author: Guillaume BUIRET, Service d’ORL et de chirurgie cervicofaciale, Centre Hospitalier de Valence, 159 Boulevard du Maréchal Juin, 26953 Valence.

Citation: Buiret G. and Chidiac F., (2021) Efficacy of Somnodent® Appliance Therapy in Moderate to Severe Obstructive Sleep Apnea Patients. J, Clinical Otorhinolaryngology, 3(3); DOI: 10.31579/2692-9562/030

Copyright: © 2021, Guillaume Buiret, This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited

Received: 04 April 2021 | Accepted: 13 July 2021 | Published: 20 August 2021

Keywords: obstructive sleep apnea, oral appliance, apnea hypopnea index

Abstract

Objective: Oral appliances are one of the treatments of obstructive sleep apnea hypopnea syndrome. The main objective of the study was to determine the efficacy of the Somnodent® oral appliance on Apnea-Hypopnea Index. The secondary objective was to measure the efficacy on other oxymetric parameters and to determine its tolerance and dropout rate. 

Methods: Efficacy results of the oral appliances based on the apnea hypopnea index, the oxygen desaturation index and the lower oxygen saturation reduction, appliance complications and dropout rates were retrospectively collected from 25 patients with a moderate to severe obstructive sleep apnea hypopnea syndrome treated by a Somnodent® oral appliance. The procedure was entirely performed by otolaryngologists.

Results: The AHI with OA was more significantly reduced in patients with a higher initial apnea hypopnea  index and a higher initial body mass index but reduction was not related to age and not proportional to degree of mandibular advancement.

The 50%-AHI reduction rate after OA was 55.6%, the AHI ≤5/h rate after OA was 22.2%. The OA significantly reduced the mean apnea-hypopnea index (-14.3/h, p<10-5). No patient had his AHI increased with oral appliance. One patient stopped using OA mainly because of nausea.

Conclusions: Somnodent® is an effective and well-tolerated treatment for moderate to severe obstructive sleep apnea hypopnea syndrome. 

 

Introduction

Obstructive sleep apnea hypopnea syndrome (OSAHS) is characterized by obstruction of the Upper Airways (UA) during sleep, causing desaturation and / or arousal [1]. In its severe form, its medium- and long-term consequences are of several orders (metabolic, cerebral, behavioural, and accidental). Of these risks, vascular risk is one of the greatest. This vascular risk is both cardiological, neurological and ophthalmological [2].In France, Oral Appliances (OAs) are one of the possible treatments of OSAHS with clinical signs such as headache, fatigue, drowsiness, nycturia, nocturnal suffocations, vascular comorbidities and road accident by falling asleep [3-7]. They are considered as a second-line treatment for severe OSAHS after refusal or discontinuation of Continuous Positive Airway Pressure (CPAP) and as a first-line treatment for moderate OSA [8]. All OAs are anchored on teeth and aim to advance the mandible towards the maxilla. The precise mechanism of OAs action remains unclear and is probably multifactorial: increase of the caliber of the upper airway by increasing the space behind the base of tongue and the soft palate, changing of configuration of the uvula and tongue, decreasing the UA resistance [9-12] and the stretching-receptors activity in the genioglossus muscle [11, 13].Different OA designs exist according to the teeth they rely on, the mechanism of advancement and/or the direction of the applied forces to advance the mandible. They are generally considered in the literature as a homogeneous group. However, considering them as a homogeneous group is debatable and very few comparisons of the efficacy and tolerance of different OA have been led [14]

Methods

Methods

  1. Population

The whole study took place in the ENT unit of our institution. 

Every patient who received an OA between 06/01/2019 and 10/31/2020 was retrospectively included. 

OA indications were:

  • a severe OSAHS (AHI≥30/h) with clinical signs such as headaches, fatigue, drowsiness, nycturia, nocturnal suffocations, vascular comorbidities and road accident by falling asleep and refusal of or intolerance to CPAP.
  • a moderate OSAHS (IAH between 15/h and 30/h) with clinical signs described above.

Patients were either directly referred to the ENT doctor by the general practitioner for a sleep exam, or by a pulmonologist, a neurologist or a cardiologist after completion of a sleep exam.

Polygraphies were performed in every patients. Interpretation was conducted using the American Association of Sleep Medicine criteria of sleep exams.

  1.  Initial ENT consultation

The initial ENT consultation, verifying the absence of OA contraindications, was fully described for another OA [15, 16].

After acceptance of patient’s refund by the French Social Security, a dental print by alginate was performed during ENT consultation and the print was sent to the OA production site  (you can see the technique on our YouTube channel: https://www.youtube.com/watch?v=d09mc7L6-HA&t=37s.

The George Gauge bite fork process is also available on our YouTube channel: https://www.youtube.com/watch?v=Zlab2Qeh2FI.

  1. OA

The OA design used in our unit was the Somnodent® custom-made titratable twin-block propulsion with cylinders OA manufactured by Somnomed laboratory (Beaucouzé, France). After OA ENT consultation delivery and follow up consultations:

The OA was delivered by an ENT doctor. Explanations about the set up and removal of the OA, the side effects, their management, and the maintenance of the OA were explained to the patient.Approximately two to four weeks after OA delivery, the patient was invited for the OA safety and efficacy assessment.

The multiparametric (subjective and objective) titration method [17] was performed: the OA advancement was titrated to approximately 50% of its maximum advancement when it is given to the patient. At each titration consultation (every two to four weeks), advancement was progressively increased by steps of 0.5 or 1 millimeter to ensure a maximum well-tolerated progress until improvement of clinical signs. A 1-mm-step was the most common step. The 0.5-mm-step was only used in case of pain induced by OA and its clinical ineffectiveness. The titration process can be seen on our YouTube channel: https://www.youtube.com/watch?v=OVwj0He4914&t=2s.

Once clinical signs are amended, the patient performed polygraphy in the same place and with the same sleep specialist as the first exam (before OA). Once the adjustment was optimized both clinically and with the sleep exam, the patient was seen regularly every six months during the first year and then on a yearly basis the following years. Follow up was conducted both by an ENT doctor in order to monitor the patient compliance, the deterioration of the OA, and the possible recurrence of OSA signs and symptoms [18], and by a dentist, in order to monitor the oral and dental impact.

  1. Primary and secondary objectives 

The main objective of the study was to determine OA efficacy on AHI.

The secondary objectives of this study were to determine rates of:

  • OA efficacy on other oxymetric parameters
  • OA tolerance
  • OA dropouts
  1. Primary and secondary outcomes

The main outcome measure was the most frequently used criteria of efficacy existing in the literature: 

  • A 50%-AHI reduction after OA
  • A residual AHI ≤5/h.
  • A <50% AHI reduction and / or residual AHI>10/h.

The secondary outcomes measures were:

  • The impact of OA on other parameters that are vitally important and less subjectively scored than the AHI: Oxygen desaturation index (ODI), lowest oxygen saturation (LOS) and time below 90% saturation (TB90S) 
  • The rates of problem of OA tolerance: not enough or too much retention, pain, gingival bleeding for example…
  • The rates and causes of OA dropout.
  1. Statistical studies

Mean comparisons were performed with Student tests and ANOVA.

Our model was built using univariate linear modeling. If necessary, multivariate linear modeling was performed with putative confounding factors.

Every test and modeling were performed with R software.

  1. Ethical consideration

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. 

For this type of study, formal consent from patients or the statement of institutional review board are not required in France.

Results

General data

Twenty-seven patients (21 males and 6 females) received an OA between 06/01/2019 and 12/31/2020. The mean age was 58.9 ± 13.9 years. The mean Body Mass Index (BMI) was 28 ± 5.1 kg/m². Ten patients (37%) stopped their CPAP treatment because of they could not tolerate it and eight (29.6%) refused it. The other characteristics of the population are reported in Table 1. The flowchart is shown in Figure 1. Twenty-two patients were finally included (21 patients with polygraphic or polysomnographic control and one patient who gave-up OA before with polygraphic or polysomnographic control). Five patients are waiting for their sleep exam control.

AHI at diagnosis significantly increased with age but not with the Body Mass Index (BMI) (p=0.0387 and 0.5621 respectively). In other words, initial AHI was significantly higher in older and but not in heavier patients.

Figure 1: Study flowchart

1.Delivery of the OA

At its delivery, OA appeared ill-adapted for three patients (12%) because of too much teeth retention leading to pain. New dental prints were immediately performed to redo another OA. 

As shown in Figure 1, five patients (18.5%) did not undergo their sleep exam control at the time of the redaction of this article.

The figure 2 presents variability of the absolute and relative advancements applied thanks to the OA. 

Figure 2: Absolute advancement (A) and relative advancement at last consultation (mm) / maximal advancement (mm) (B)

2.Effect of OA on AHI

The mean AHI reduction with OA was 14.3/h ± 7.2 (p<10-5). 55.6% of patients had a >50% AHI reduction after OA. The individual absolute and relative reductions of AHI are presented in Figure 3. No patient had an AHI increase with OA (Figure 3).

The other outcomes are shown in Table 1. 

Table 1: population characteristics. SE: standard-error. AHI: Apnea-Hypopnea Index. OSAHS: obstructive sleep apnea hypopnea syndrome. OA: Oral appliance. ODI: Oxygen Desaturation Index. LOS: Lowest Oxygen Saturation. TB90%S: Time below 90% saturationRelative advancement = advancement at the last consultation (mm) / maximal advancement (mm)
Figure 3: Apnea Hypopnea Index (AHI) reduction. 

A: Absolute AHI change (/h). A positive absolute change means decrease of the AHI after oral appliance; a negative absolute change is increase of the AHI after oral appliance.

B: Relative change of AHI (%) = 100 x (final AHI-initial AHI) / initial AHI. A positive relative change means a decrease of the AHI after oral appliance, a negative relative change indicates an increase of the IAH after oral appliance 

 

OA: Oral Appliance, OSAHS: Obstructive Sleep Apnea Hypopnea Syndrome

Table 2 presents changes of OSAHS severity classes.

The absolute and relative changes of AHI per patient are shown in Figure 4. No main characteristics of the patients (age, BMI, initial AHI) had a significant impact on the AHI absolute and relative reductionThe absolute and relative changes of AHI
The absolute and relative changes of AH per patient are shown in Figure 4. No main characteristics of the patients (age, BMI, initial AHI) had a significant impact on the AHI absolute and relative reduction.

Figure 4: A: Absolute Apnea Hypopnea Index (AHI) change according to the age B: Relative AHI change according to the age C: Absolute AHI change according to the initial body mass index (BMI) D: Relative AHI change according to the initial BMI E: Absolute AHI change according to the initial AHI F: Relative AHI change according to the initial AHI G: Absolute AHI change according to the absolute advancementH: Relative AHI change according to the absolute advancement I: Absolute AHI change according to the relative advancement J: Relative AHI change according to relative advancement
  1. Effect of OA on other vital parameters

Other outcomes are shown in Table 1. 

The mean ODI significantly decreased with OA: -9.3 events/h, p=0.021. The mean LOS did not significantly increase with OA: +0.13%, p=0.9588. So did the mean TB90S (-13.2/h, p=0.0967).

  1. Compliance and adherence to the OA

Tolerance difficulty (nausea) was reported by one patient (4%) from the second ENT consultation leading to the discontinuation of the use of OA. 

 

Discussion

 Our study has demonstrated that Somnodent® OA allowed a significant efficacy on AHI (AHI mean decrease of 14.3/h ± 7.2 and a 50%-AHI reduction rate of 55.4%) and on ODI with a mean advancement of 5.3mm.  The efficacy seemed very close to the main studies of Somnodent® OA where the mean AHI decrease was -16/h for Mehta et al. [19] (28 patients), -15/h for Gotsopoulos et al. [20] (73 patients), -13/h (23 patients) for Pitsis et al, -11.8/h for Ng et al. [11] (10 patients). The mean AHI reduction was also similar to our experience in Narval® OA [15] (-14.9/h ± 11.9/h and a 50%-AHI reduction rate of 65.2%).

Criteria for OAs efficacy vary according to the studies. The main outcome that we chose to evaluate OA efficacy (AHI decrease ability) is debatable because of its inter- and intra-observer measurement variability and the fact that it is not a clinical parameter. But the impact on AHI is also present on ODI and LOS (more objective but less sensitive parameters) and those parameters are also often retained [6, 15, 16, 20-24].

The OA tolerance was excellent, since only one patient (3.7%) could not tolerate it, a finding that was consistent with the rates in the literature [7, 25].

Although our study is not comparative, compliance seemed better than the rates reported in the literature in CPAP follow up studies [26, 27] or in studies comparing CPAP and OA [28-30] and comaprable than our experience in Narval® OA [15].

Our study may be criticized because it is retrospective with many patients that did not have a sleep exam control (18.5%). It must be considered as an evaluation of experience, first step before a comparative study between Narval® OA from Resmed laboratory (Saint-Genis-Laval, France), necessary to calculate the sample size. Indeed our ENT unit has been using the Narval® OA from for many years. It is an efficient design that is well tolerated by patients [671516233132]. We first used the Somnodent® OA to treat patients without teeth and later for any patients with an apparent similar efficacy and tolerance. After having evaluated the efficacy and tolerance of each OA, the next step will be to compare one to another.

Conclusion

In this study, Somnodent® OA led to a significant AHI and ODI reduction in moderate to severe OSAHS patients. Comparative studies should be performed to determine if a particular type of OA is more effective or better tolerated than another.

Conflict of Interest

Authors declare that they have no conflict of interest. 

Funding

 The work was not supported by any fund.

Ethical approval

All applicable international, national, and/or institutional guidelines for the care and use of animals were followed. All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards."

Informed consent

 For this type of study, formal consent is not required.

Acknowledgement

To Dr. B Navailles, MD, former head of the ENT Head and Neck surgery department, for developing oral appliances and for the interest he has elicited in all members of his team. RIP.

References

Clearly Auctoresonline and particularly Psychology and Mental Health Care Journal is dedicated to improving health care services for individuals and populations. The editorial boards' ability to efficiently recognize and share the global importance of health literacy with a variety of stakeholders. Auctoresonline publishing platform can be used to facilitate of optimal client-based services and should be added to health care professionals' repertoire of evidence-based health care resources.

img

Virginia E. Koenig

Journal of Clinical Cardiology and Cardiovascular Intervention The submission and review process was adequate. However I think that the publication total value should have been enlightened in early fases. Thank you for all.

img

Delcio G Silva Junior

Journal of Women Health Care and Issues By the present mail, I want to say thank to you and tour colleagues for facilitating my published article. Specially thank you for the peer review process, support from the editorial office. I appreciate positively the quality of your journal.

img

Ziemlé Clément Méda

Journal of Clinical Research and Reports I would be very delighted to submit my testimonial regarding the reviewer board and the editorial office. The reviewer board were accurate and helpful regarding any modifications for my manuscript. And the editorial office were very helpful and supportive in contacting and monitoring with any update and offering help. It was my pleasure to contribute with your promising Journal and I am looking forward for more collaboration.

img

Mina Sherif Soliman Georgy

We would like to thank the Journal of Thoracic Disease and Cardiothoracic Surgery because of the services they provided us for our articles. The peer-review process was done in a very excellent time manner, and the opinions of the reviewers helped us to improve our manuscript further. The editorial office had an outstanding correspondence with us and guided us in many ways. During a hard time of the pandemic that is affecting every one of us tremendously, the editorial office helped us make everything easier for publishing scientific work. Hope for a more scientific relationship with your Journal.

img

Layla Shojaie

The peer-review process which consisted high quality queries on the paper. I did answer six reviewers’ questions and comments before the paper was accepted. The support from the editorial office is excellent.

img

Sing-yung Wu

Journal of Neuroscience and Neurological Surgery. I had the experience of publishing a research article recently. The whole process was simple from submission to publication. The reviewers made specific and valuable recommendations and corrections that improved the quality of my publication. I strongly recommend this Journal.

img

Orlando Villarreal

Dr. Katarzyna Byczkowska My testimonial covering: "The peer review process is quick and effective. The support from the editorial office is very professional and friendly. Quality of the Clinical Cardiology and Cardiovascular Interventions is scientific and publishes ground-breaking research on cardiology that is useful for other professionals in the field.

img

Katarzyna Byczkowska

Thank you most sincerely, with regard to the support you have given in relation to the reviewing process and the processing of my article entitled "Large Cell Neuroendocrine Carcinoma of The Prostate Gland: A Review and Update" for publication in your esteemed Journal, Journal of Cancer Research and Cellular Therapeutics". The editorial team has been very supportive.

img

Anthony Kodzo-Grey Venyo

Testimony of Journal of Clinical Otorhinolaryngology: work with your Reviews has been a educational and constructive experience. The editorial office were very helpful and supportive. It was a pleasure to contribute to your Journal.

img

Pedro Marques Gomes

Dr. Bernard Terkimbi Utoo, I am happy to publish my scientific work in Journal of Women Health Care and Issues (JWHCI). The manuscript submission was seamless and peer review process was top notch. I was amazed that 4 reviewers worked on the manuscript which made it a highly technical, standard and excellent quality paper. I appreciate the format and consideration for the APC as well as the speed of publication. It is my pleasure to continue with this scientific relationship with the esteem JWHCI.

img

Bernard Terkimbi Utoo

This is an acknowledgment for peer reviewers, editorial board of Journal of Clinical Research and Reports. They show a lot of consideration for us as publishers for our research article “Evaluation of the different factors associated with side effects of COVID-19 vaccination on medical students, Mutah university, Al-Karak, Jordan”, in a very professional and easy way. This journal is one of outstanding medical journal.

img

Prof Sherif W Mansour

Dear Hao Jiang, to Journal of Nutrition and Food Processing We greatly appreciate the efficient, professional and rapid processing of our paper by your team. If there is anything else we should do, please do not hesitate to let us know. On behalf of my co-authors, we would like to express our great appreciation to editor and reviewers.

img

Hao Jiang

As an author who has recently published in the journal "Brain and Neurological Disorders". I am delighted to provide a testimonial on the peer review process, editorial office support, and the overall quality of the journal. The peer review process at Brain and Neurological Disorders is rigorous and meticulous, ensuring that only high-quality, evidence-based research is published. The reviewers are experts in their fields, and their comments and suggestions were constructive and helped improve the quality of my manuscript. The review process was timely and efficient, with clear communication from the editorial office at each stage. The support from the editorial office was exceptional throughout the entire process. The editorial staff was responsive, professional, and always willing to help. They provided valuable guidance on formatting, structure, and ethical considerations, making the submission process seamless. Moreover, they kept me informed about the status of my manuscript and provided timely updates, which made the process less stressful. The journal Brain and Neurological Disorders is of the highest quality, with a strong focus on publishing cutting-edge research in the field of neurology. The articles published in this journal are well-researched, rigorously peer-reviewed, and written by experts in the field. The journal maintains high standards, ensuring that readers are provided with the most up-to-date and reliable information on brain and neurological disorders. In conclusion, I had a wonderful experience publishing in Brain and Neurological Disorders. The peer review process was thorough, the editorial office provided exceptional support, and the journal's quality is second to none. I would highly recommend this journal to any researcher working in the field of neurology and brain disorders.

img

Dr Shiming Tang

Dear Agrippa Hilda, Journal of Neuroscience and Neurological Surgery, Editorial Coordinator, I trust this message finds you well. I want to extend my appreciation for considering my article for publication in your esteemed journal. I am pleased to provide a testimonial regarding the peer review process and the support received from your editorial office. The peer review process for my paper was carried out in a highly professional and thorough manner. The feedback and comments provided by the authors were constructive and very useful in improving the quality of the manuscript. This rigorous assessment process undoubtedly contributes to the high standards maintained by your journal.

img

Raed Mualem

International Journal of Clinical Case Reports and Reviews. I strongly recommend to consider submitting your work to this high-quality journal. The support and availability of the Editorial staff is outstanding and the review process was both efficient and rigorous.

img

Andreas Filippaios

Thank you very much for publishing my Research Article titled “Comparing Treatment Outcome Of Allergic Rhinitis Patients After Using Fluticasone Nasal Spray And Nasal Douching" in the Journal of Clinical Otorhinolaryngology. As Medical Professionals we are immensely benefited from study of various informative Articles and Papers published in this high quality Journal. I look forward to enriching my knowledge by regular study of the Journal and contribute my future work in the field of ENT through the Journal for use by the medical fraternity. The support from the Editorial office was excellent and very prompt. I also welcome the comments received from the readers of my Research Article.

img

Dr Suramya Dhamija

Dear Erica Kelsey, Editorial Coordinator of Cancer Research and Cellular Therapeutics Our team is very satisfied with the processing of our paper by your journal. That was fast, efficient, rigorous, but without unnecessary complications. We appreciated the very short time between the submission of the paper and its publication on line on your site.

img

Bruno Chauffert

I am very glad to say that the peer review process is very successful and fast and support from the Editorial Office. Therefore, I would like to continue our scientific relationship for a long time. And I especially thank you for your kindly attention towards my article. Have a good day!

img

Baheci Selen

"We recently published an article entitled “Influence of beta-Cyclodextrins upon the Degradation of Carbofuran Derivatives under Alkaline Conditions" in the Journal of “Pesticides and Biofertilizers” to show that the cyclodextrins protect the carbamates increasing their half-life time in the presence of basic conditions This will be very helpful to understand carbofuran behaviour in the analytical, agro-environmental and food areas. We greatly appreciated the interaction with the editor and the editorial team; we were particularly well accompanied during the course of the revision process, since all various steps towards publication were short and without delay".

img

Jesus Simal-Gandara

I would like to express my gratitude towards you process of article review and submission. I found this to be very fair and expedient. Your follow up has been excellent. I have many publications in national and international journal and your process has been one of the best so far. Keep up the great work.

img

Douglas Miyazaki

We are grateful for this opportunity to provide a glowing recommendation to the Journal of Psychiatry and Psychotherapy. We found that the editorial team were very supportive, helpful, kept us abreast of timelines and over all very professional in nature. The peer review process was rigorous, efficient and constructive that really enhanced our article submission. The experience with this journal remains one of our best ever and we look forward to providing future submissions in the near future.

img

Dr Griffith

I am very pleased to serve as EBM of the journal, I hope many years of my experience in stem cells can help the journal from one way or another. As we know, stem cells hold great potential for regenerative medicine, which are mostly used to promote the repair response of diseased, dysfunctional or injured tissue using stem cells or their derivatives. I think Stem Cell Research and Therapeutics International is a great platform to publish and share the understanding towards the biology and translational or clinical application of stem cells.

img

Dr Tong Ming Liu

I would like to give my testimony in the support I have got by the peer review process and to support the editorial office where they were of asset to support young author like me to be encouraged to publish their work in your respected journal and globalize and share knowledge across the globe. I really give my great gratitude to your journal and the peer review including the editorial office.

img

Husain Taha Radhi

I am delighted to publish our manuscript entitled "A Perspective on Cocaine Induced Stroke - Its Mechanisms and Management" in the Journal of Neuroscience and Neurological Surgery. The peer review process, support from the editorial office, and quality of the journal are excellent. The manuscripts published are of high quality and of excellent scientific value. I recommend this journal very much to colleagues.

img

S Munshi

Dr.Tania Muñoz, My experience as researcher and author of a review article in The Journal Clinical Cardiology and Interventions has been very enriching and stimulating. The editorial team is excellent, performs its work with absolute responsibility and delivery. They are proactive, dynamic and receptive to all proposals. Supporting at all times the vast universe of authors who choose them as an option for publication. The team of review specialists, members of the editorial board, are brilliant professionals, with remarkable performance in medical research and scientific methodology. Together they form a frontline team that consolidates the JCCI as a magnificent option for the publication and review of high-level medical articles and broad collective interest. I am honored to be able to share my review article and open to receive all your comments.

img

Tania Munoz

“The peer review process of JPMHC is quick and effective. Authors are benefited by good and professional reviewers with huge experience in the field of psychology and mental health. The support from the editorial office is very professional. People to contact to are friendly and happy to help and assist any query authors might have. Quality of the Journal is scientific and publishes ground-breaking research on mental health that is useful for other professionals in the field”.

img

George Varvatsoulias

Dear editorial department: On behalf of our team, I hereby certify the reliability and superiority of the International Journal of Clinical Case Reports and Reviews in the peer review process, editorial support, and journal quality. Firstly, the peer review process of the International Journal of Clinical Case Reports and Reviews is rigorous, fair, transparent, fast, and of high quality. The editorial department invites experts from relevant fields as anonymous reviewers to review all submitted manuscripts. These experts have rich academic backgrounds and experience, and can accurately evaluate the academic quality, originality, and suitability of manuscripts. The editorial department is committed to ensuring the rigor of the peer review process, while also making every effort to ensure a fast review cycle to meet the needs of authors and the academic community. Secondly, the editorial team of the International Journal of Clinical Case Reports and Reviews is composed of a group of senior scholars and professionals with rich experience and professional knowledge in related fields. The editorial department is committed to assisting authors in improving their manuscripts, ensuring their academic accuracy, clarity, and completeness. Editors actively collaborate with authors, providing useful suggestions and feedback to promote the improvement and development of the manuscript. We believe that the support of the editorial department is one of the key factors in ensuring the quality of the journal. Finally, the International Journal of Clinical Case Reports and Reviews is renowned for its high- quality articles and strict academic standards. The editorial department is committed to publishing innovative and academically valuable research results to promote the development and progress of related fields. The International Journal of Clinical Case Reports and Reviews is reasonably priced and ensures excellent service and quality ratio, allowing authors to obtain high-level academic publishing opportunities in an affordable manner. I hereby solemnly declare that the International Journal of Clinical Case Reports and Reviews has a high level of credibility and superiority in terms of peer review process, editorial support, reasonable fees, and journal quality. Sincerely, Rui Tao.

img

Rui Tao

Clinical Cardiology and Cardiovascular Interventions I testity the covering of the peer review process, support from the editorial office, and quality of the journal.

img

Khurram Arshad

Clinical Cardiology and Cardiovascular Interventions, we deeply appreciate the interest shown in our work and its publication. It has been a true pleasure to collaborate with you. The peer review process, as well as the support provided by the editorial office, have been exceptional, and the quality of the journal is very high, which was a determining factor in our decision to publish with you.

img

Gomez Barriga Maria Dolores

The peer reviewers process is quick and effective, the supports from editorial office is excellent, the quality of journal is high. I would like to collabroate with Internatioanl journal of Clinical Case Reports and Reviews journal clinically in the future time.

img

Lin Shaw Chin

Clinical Cardiology and Cardiovascular Interventions, I would like to express my sincerest gratitude for the trust placed in our team for the publication in your journal. It has been a true pleasure to collaborate with you on this project. I am pleased to inform you that both the peer review process and the attention from the editorial coordination have been excellent. Your team has worked with dedication and professionalism to ensure that your publication meets the highest standards of quality. We are confident that this collaboration will result in mutual success, and we are eager to see the fruits of this shared effort.

img

Maria Dolores Gomez Barriga