Dispersive Liquid - Liquid Micro extraction Technique Coupled with High Performance Liquid Chromatography Diode Array Detector for the Determination of Sulfonamides in Water Samples

Research Article

Dispersive Liquid - Liquid Micro extraction Technique Coupled with High Performance Liquid Chromatography Diode Array Detector for the Determination of Sulfonamides in Water Samples

  • Eskinder Teklu Bekele 1,2*
  • Dube Simiso 1
  • Nindi Mathew Muzi 3

*Corresponding Author: Eskinder Teklu Bekele, Department of Chemistry, College of Science Engineering and Technology, Science Campus, University of South Africa, Corner Christian de Wet and Pioneer Avenue Florida Park, Roodepoort 1709, Gauteng, South Africa.

Citation: : Eskinder T. Bekele, Dube Simiso, and Nindi M. Muzi (2023), Dispersive Liquid - Liquid Micro extraction Technique Coupled with High Performance Liquid Chromatography Diode Array Detector for the Determination of Sulfonamides in Water Samples. J. Nutrition and Food Processing, 6(1); DOI:10.31579/2637-8914/115

Copyright: © 2023 Eskinder Teklu Bekele, This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Received: 10 January 2023 | Accepted: 17 February 2023 | Published: 20 March 2023

Keywords: dispersive liquid-liquid microextraction (DLLME); sulfonamides (SAs); HPLC-DAD

Abstract

A dispersive liquid-liquid microextraction combined with high-performance liquid chromatography with diode array detection (DLLME-HPLC-DAD) has been  developed for the determination of 15 sulfonamides in water samples. The effects of experimental   parameters on the performance of the method such as the type and volume of extraction and disperser solvents, pH of the solution, effect of salt, and centrifugation time were evaluated, and optimum conditions were established. Under these optimum conditions, linearity was found in the range of 2 - 1 000 μg L-1 with regression coefficients better than 0.9990. The limit of detection (LOD) and limit of quantification (LOQ) values were in the range of 0.6 - 7.8 μg L-1 and 1.5 - 8.3 µg L-1, respectively. Intra-day and inter-day precision   ranged from 1.5 - 9.7% and from 3.1 - 9.9%, respectively. The accuracy of the method were also evaluated in  samples from three different sources of water (river water, groundwater,  and tap water) were found to  range from 70.5 to 103.9% with %RSD values ranging from 0.5 to 9.8%, with the exception of sulfaguanidine in river water and groundwater where the percentage recovery ranged from 54.3 - 84.9% at the two higher concentration levels (100 and 400 µg L-1).Furthermore, the effect of sample matrix on the extraction of target compounds was evaluated using the relative recovery (%RR) of the samples from three different water sources relative to the ultrahigh purity (UHP) water sample spiked with target analytes, and found in  the range of  70.3 - 106.1%.The proposed method was successfully applied to wastewater treatment plant samples and  Sulfisoxazole (SSO) was detected in the influent samples in the range of 14.8 - 17.8 µg L-1 with corresponding %RSD, (n = 5) value of 1.6% to 1.9%.The results indicated that the proposed method is effective for the extraction and determination of the sulfonamides in water samples.

1.Introduction

Currently,  the use of antibiotics in human and veterinary medicine has become a common practice that has led to the gradual accumulation of antibiotics in the environment, which enter through sources of contamination such as wastewater, landfills, urban sites, as well as industrial and hospital effluents.(Manzetti and Ghisi 2014) Furthermore, due to the limited knowledge of the input, fate, and effects of most antibiotics in the environment, there is no regulation on the levels of these compounds in environmental matrices, including water systems, sediment, and soil  (Kümmerer 2009)However, concerns about the occurrence, transport, and fate of veterinary drugs in the environment have been ever increasing in recent years, and several initiatives  have been launched to establish or strengthen the monitoring systems.(Seifrtová, Pena et al. 2008Pan, Qiang et al. 2011Iglesias, Nebot et al. 2013
Sulfonamides may enter natural systems following different paths, via  discharge from wastewater treatment  plants (WWTPs), leakage  from septic systems and agricultural waste storage facilities as well as the application of human and agricultural waste to land to supplement fertilisers.(García-Galán, Silvia Díaz-Cruz et al. 2008Kümmerer 2009Baran, Adamek et al. 2011) Sulfonamides are fairly soluble in water and their ionisation is highly dependent on the pH of the matrix (structures, KOW. and pKa values are shown in Table S1, supplemental material). Thus, after disposal in soils they may enter surface runoff or be leached into the groundwater. Furthermore, these compounds may not be completely metabolised and a high proportion of them are possibly excreted directly in faeces (digested wastes) and urine. Therefore, both the parent compounds and their metabolites can be released directly to the environment.(Białk-Bielińska, Stolte et al. 2011) It is therefore important to develop simple, efficient, and eco-friendly analytical methods for the monitoring of these drug residues. Various methods have been developed for the determination of SA residues, such as gas chromatography-mass spectrometry (GC-MS),(Chiavarino, Crestoni et al. 1998) capillary electrophoresis with ultraviolet detection (CE-UV),(Wen, Li et al. 2011)and high-performance liquid chromatography (HPLC) with ultraviolet (UV),(Chitescu, Nicolau et al. 2011) fluorescence detection (FD),(Mor, Sahindokuyucu Kocasari et al. 2012Arroyo-Manzanares, Gámiz-Gracia et al. 2014) MS or MS/MS,(Senta, Terzić et al. 2008García-Galán, Díaz-Cruz et al. 2010Tölgyesi, Berky et al. 2013) and electrochemical methods.(Fotouhi, Fatollahzadeh et al. 2012). 
Since sulfonamides are found at very low (µg L-1) levels and in most cases more than one type of compound (parent and metabolites) would be determined, efficient sample preparation (extraction, enrichment, and/or clean up) techniques are mandatory prior to instrumental analysis. Liquid-liquid extraction,(Koesukwiwat, Jayanta et al. 2007) solid-phase extraction,(García-Galán, Díaz-Cruz et al. 2010)solid-phase microextraction, (Lu, Chen et al. 2007) and pressurised liquid extraction(Yu, Tao et al. 2011) have thus been widely used as sample preparation methods for the extraction of sulfonamides from environmental and food matrices. Solid-phase microextraction (SPME) offers simpler steps, compatible with GC and LC separation techniques, and it is almost solvent-free, but it has inherent drawbacks in that the fibre can break easily, the fibre coating can be damaged, and the fibre assembly is relatively expensive.(Schmidt and Podmore 2015) On the other hand, pressurised liquid extraction (PLE) requires expensive instrumentation, and it is less selective.(Yu, Tao et al. 2011
In recent years, liquid-phase microextraction (LPME) techniques have become widely applied techniques in analytical research areas,(Sarafraz-Yazdi and Amiri 2010Asensio-Ramos, Ravelo-Pérez et al. 2011Prosen 2014) due to their consumption of very small volumes (µL) of extraction solvent; one of the objectives of green chemistry is addressed by minimising the amount of solvent usage (miniaturisation) in the analytical process.
Among the above-mentioned recently developed sample preparation techniques, dispersive liquid-liquid microextraction (DLLME) is of particular interest due to the fact that the technique has several advantages, including simplicity of operation, rapidity, low cost, high recovery, and enrichment factors.(Rezaee, Assadi et al. 2006Yan and Wang 2013). In DLLME, extraction is carried out between the sample and a cloud of fine extractant droplets formed when the mixture of extraction solvent, which is water immiscible, and the disperser solvent, which is miscible with both aqueous solvents and extraction solvents, is injected rapidly into an aqueous sample containing the analytes of interest.(Yan and Wang 2013).  A very high contact area is generated between the aqueous phase and the extraction solvent; thus, analytes are rapidly extracted into the extraction phase.(Yan and Wang 2013). Therefore, analytes have to be in their neutral form for them to efficiently partition between the aqueous phase and the hydrophobic organic phase (extraction solvent). This explains why the technique was initially applied to neutral or non-polar organic compounds such as PAH which have larger octanol-water partition coefficient (KOW) values.  However, the application of   DLLME for polar analytes poses a challenge, since they will rather remain in the aqueous phase than be extracted into the organic solvent. These analytes are soluble in water mainly in their ionised form; therefore, appropriate adjustment of sample pH is necessary to suppress ionisation of target analytes. Furthermore, by using an appropriate amount of salt to induce a “salting-out effect”, the extractability of polar analytes can be improved as their solubility in the aqueous phase is reduced. Compounds with a wider range of polarity like sulfonamides present yet another dimension of complexity. As a result of such challenges, most of the reported DLLME methods often focused on the less polar group of sulfonamide compounds(Wen, Li et al. 2011Xu, Su et al. 2011Li, Li et al. 2016Wang, Li et al. 2022) with  a few exceptions  occasionally being reported for some polar group of sulfonamides such as Herrera-Herrera et al.(Herrera-Herrera, Hernández-Borges et al. 2013) in 2013 reported for the determination of 11 sulfonamides together with 14 fluoroquinolones using DLLME coupled with UHPLC-DAD. However, most of the reported methods require longer extraction times or use of advanced instrument. Furthermore, in some cases a higher volume of solvents and detection limits were reported. Therefore, it is desirable to develop a fast, cost effective and sensitive analytical method for simultaneous extraction and determination of wide number of sulfonamides with diverse polarity ranges.
Thus, the aim of the current study was to develop a fast, cost effective and sensitive DLLME method coupled with HPLC –DAD for the simultaneous extraction and determination of 15 sulfonamide compounds in diverse water samples. The effects of various parameters on the extraction efficiencies of target compounds such as the type and volume of extraction and disperser solvents, pH of the solution, effect of salt addition and centrifuge time for DLLME procedure were   evaluated and optimum conditions were established. The performance of the developed method was validated and applied to wastewater samples. Results revealed that the proposed method is effective to be used for routine monitoring of sulfonamides and related drugs in diverse water samples

2. Methods and Materials

2.1. Standards and chemicals
Antibiotic standards included sulfaguanidine (SGD), sulfanilamides (SAM), sulfathiazole (STZ), sulfacetamides (SAA), sulfamethizole (SMT), sulfamethoxazole (SMX), sulfasalazine (SSA), sulfamonomethoxine (SMM), sulfaquinoxaline (SQX), sulfamerazine (SMR), sulfapyridine (SPY), sulfadiazine (SDZ), sulfabenzamide (SBZ), sulfachloropyridine (SCP), and sulfisoxazole (SSO), and were purchased from Sigma-Aldrich (Steinheim, Germany). All standards had purity higher than 98%. Methanol, acetone, and formic acid were obtained from Sigma - Aldrich (Steinheim, Germany). Acetonitrile (MeCN) was purchased from ROMIL Ltd. (Waterbeach, Cambridge, UK). Sodium   hydroxide (NaOH) was supplied by Merck (Darmstadt, Germany). Trichloromethane, dichloromethane, 1,2-dichloroethane and 1,1,2,2-tetrachloroethane were of HPLC grade and were purchased from Sigma-Aldrich (Steinheim, Germany). Ultrahigh purity (UHP) water (resistivity, 18.2 MΩ∙cm at 25 ℃) was generated using the Milli-Q® system (Millipore, Billerica, MA, USA). 
2.2. Sampling of tap water, groundwater, river water and wastewaters from wastewater treatment plant (influents and effluent)
For the purposes of validating the proposed method, four different types of water samples were collected using 2.5 L sampling bottles: Tap water I sample from Pretoria, South Africa; tap water II sample from Florida, South Africa; river water sample from Suikerbosrand River (Gauteng, South Africa); and groundwater sample from Florida, South Africa.  Wastewater   treatment plant samples (influent and effluent) were collected from the Daspoort Wastewater Treatment Plant (WWTP) located at latitude 25.73450 south and longitude 28.17810   East coordinates in Pretoria, Gauteng Province, South Africa. These samples were collected from three different locations: -i) before entry into the treatment plant (Influent I); ii) before entry into the treatment plant from a different location (influent (II); and iii) after treatment at the outlet (effluent). The grab sampling method was used, and the samples were collected into amber glass bottles, which had previously been acid-washed and rinsed with UHP water and then flushed with wastewater before collection. Each sample was collected in triplicate into 2.5 L bottles. The water samples were filtered through Whatman filter paper and 0.45 µm nylon filters, respectively, and then stored in amber glass bottles in a refrigerator at 4 ℃ until analysis. 
2.3. Instrumentation
An Agilent 1260 series high-performance liquid chromatographic system (Agilent Technologies, Waldbronn, Germany) was used for all separations. The HPLC consisted of a binary pump, vacuum degasser, thermostatted column compartment, auto-sampler, diode array detector (DAD), and fluorescence detector (FLD). Data acquisition was achieved using the Agilent ChemStation (version 1.9.0) software. Chromatographic separations were carried out using ZORBAX Eclipse Plus C18 column (100 mm x 4.6 mm, 3.5 µm) from Agilent Technologies, Inc. (Santa Clara, CA, USA). A vortex mixer (VELP Scientifica, Usmate Velate (MB), Italy), centrifuge (Thermo Electron Corporation, Massachusetts, USA) and a micro balance (Mettler Toledo XP6U) were used for sample preparation. A Hamilton® 500 µL syringe (Hamilton Bonaduz AG, Switzerland) was used for withdrawing the settled phase in the DLLME procedure and nitrogen gas was used for drying the samples.
2.4. Preparation of sulfonamide standard solutions
Stock standard solutions (1 000 mg L-1) for each compound were prepared by dissolving10 mg of accurately weighed standard of each compound in 10 mL of a mixture of methanol and ultrahigh purity water (1:1; v/v). Appropriate dilution of these stock solutions with methanol-ultrahigh purity water (1:1; v/v) was used to prepare various concentrations of working solutions. All standard solutions were protected from light by covering the sample vials with aluminium foil and were kept at 4 ℃ until use.
2.5. Optimisation of chromatographic conditions for the separation of 15 sulfonamide compounds
A chromatographic method for the separation of 15 sulfonamides was first developed and optimised. To investigate the conditions for optimum peak shape and adequate resolution in the separation of target analytes, both ultrahigh purity (UHP) water and acidified ultrahigh purity water were evaluated as a solvent A and organic solvents (acetonitrile and methanol) as a solvent B. The effect of mobile phase flow rate was also evaluated in the range of 0.3 - 2 mL min-1 at 0.2 mL min-1 interval. Both isocratic and gradient elution modes were investigated Furthermore, the column oven temperature was also optimised in the range of 25 - 45 ºC and compounds were monitored using DAD at different wavelengths (260, 265, 270 and 280 nm). 
2.6. Optimisation of DLLME conditions
In DLLME, the factors that can affect the efficiency of extraction include type and volume of extraction and disperser solvent, the pH of the solution, salting-out effect and the extraction time. A disperser solvent must disperse the extraction solvent (form fine droplets) into the aqueous solution while the volume of extraction and disperser solvents would have a significant effect on the enrichment factor (EF). Therefore, it is important to optimise all the factors in order to obtain good performance.
2.6.1. Evaluation of extraction solvent
In DLLME  it is very important  to select an appropriate extraction solvent that is capable of extracting  the target compound, has a low solubility in water, good chromatographic behaviour (i.e. should not absorb  at the wavelength of target analytes), has  a higher   density than water, and is able to form two phases in the presence of a disperser solvent.(Rezaee, Assadi et al. 2006Schmidt and Podmore 2015). Based on the above criteria, dichloromethane, CH2Cl2 (1.32 g mL-1), trichloromethane, CHCl3 (1.47 g mL-1), 1,2-dichloroethane, C2H4Cl2 (1.25 g mL-1), and 1,1,2,2-tetrachloroethane, C2H2Cl4 (1.59 g mL-1) were selected as potential solvents for this study. The extraction solvents were   evaluated at a volume of 600 µL and using 1 mL of MeCN as the disperser solvent; these are conditions adapted from previous reports in the literature.(Rezaee, Assadi et al. 2006Wen, Li et al. 2011Xu, Su et al. 2011Yan and Wang 2013Li, Li et al. 2016).
2.62. Effect of the volume of extraction solvent
The volume of the extraction solvent has an impact on the enrichment factors. If very low volumes are used, the extraction solvent may not be distributed well in the aqueous phase resulting in low enrichment factors. On the other hand, high volumes of extraction solvent may lower the optimum ratio of the disperser solvent to the extraction solvent volume. This consequently reduces the amount of droplets formed which are necessary for the extraction process thereby lowering the enrichment factors.(Saraji, Khalili Boroujeni et al. 2011) Therefore, the effect of extraction solvent volume on the enrichment factor was investigated over a range of 200 - 1 000 μL, while all other experimental parameters were maintained constant. 
2.6.3. Evaluation of disperser solvent 
Evaluation of an appropriate disperser solvent is important since it can affect the enrichment factors of target analytes. Disperser solvents must fulfil the condition of being miscible with both water and the extraction solvents.(Rezaee, Assadi et al. 2006
In this study, methanol, acetonitrile, ethanol and acetone were selected as potential disperser solvents and their effect on the enrichment factor was evaluated using 1 mL of each solvent and 600 μL of dichloromethane as the extraction solvent.
2.6.4. Effect of the volume of disperser solvent
To determine the optimum volume of the disperser solvent, various volumes (200 - 1 200 μL) of acetonitrile containing a fixed volume of dichloromethane (400 μL) were evaluated. 
2.6.5. Effect of the solution pH 
In principle, DLLME is most appropriate for extracting neutral or non-charged compounds from aqueous to organic phase, thus in a way, making it a limited technique. Therefore, pH control becomes essential to extend is applicability to ionic and polar compounds such as sulfonamides. Theoretically, SAs are positively charged in acidic medium at two pH units below their acidic pKa values, neutral at pH between pKa1 and   pKa2, and negatively   charged at 3.3 pH units above their basic pKa2 values. The pKa1 values of target compounds are in the range of 0.9 - 2.5 and their pKa2 values are in the range of 5.3 - 11.3. One of the challenges when dealing with analytes of diverse polarity is to find a suitable pH in which the majority of target analytes can be    extracted in their neutral form. Therefore, the selected pH must be adequate for the extraction of all analytes. Thus, the effect of the solution pH  was   evaluated by varying it in the range of 2.5 - 7.5 using HCl (0.1 M) and/or NaOH (0.1 M) and keeping the other conditions constant. 
2.6.6 Effect of salt addition
The addition of salt into the sample solution decreases the solubility of the analytes, thus resulting in the transfer of analytes into the organic phase. Therefore, the effect of salt on the enrichment factors of target compounds was tested by adding NaCl in the range of 0 - 8 % (w/v), while keeping all the other experimental conditions constant(Xu, Su et al. 2011). 
2.6.7 Effect of centrifugation time
Adequate centrifugation time is important for the formation of phase separation. A very long centrifugation should be avoided because a centrifuge generates heat as it rotates, which can destabilise phase separation. Therefore, in this study, the effect of centrifugation time was evaluated in the range of 1 - 10 min. 
2.7. Dispersive liquid-liquid microextraction procedure for water samples
Into 15 mL screw cap centrifuge tubes, 5 mL of ultrahigh purity (UHP) water, with the pH previously adjusted to 3.5 using 0.1 M HCl that contained 0.10 g (2%; w/v) NaCl, was transferred, and spiked with a mixture of SAs at a concentration of 100 μg L-1. Thereafter, the mixture of extractant and disperser (400 µL and 600 µL, respectively) was    rapidly    injected into the aqueous solution containing the sample using an aseptic disposable syringe. The mixture was then vortexed for 30 s to ensure complete dispersion and to facilitate the extraction. The dispersed fine particles of the extraction phase, which had settled at the bottom of a screw cap centrifuge tube, were transferred to a 1.5 mL HPLC vial (Agilent, Germany) through a 400-µL vial insert using a Hamilton 500-µL microsyringe and dried under a gentle stream of nitrogen gas. Thereafter, the remaining residue was   reconstituted with 100 µL of the mobile phase and subjected to chromatographic analysis.
2.8. Chromatographic conditions for the separation of 15 sulfonamide compounds
The chromatographic conditions for the separation of 15 SAs were as follows:  a binary mobile phase comprising of solvent A (0.1% formic acid at pH 2.73) and solvent B (acetonitrile) with a gradient elution of 10% B (0 - 1 min), which was gradually increased from 10% to 40

3. Results and Discussion

For chromatic separation of sulfonamide compounds, optimum resolution and peak shape were obtained at a flow rate of 1.8 mL min-1. The best peak shape and satisfactory resolution of the target compounds were achieved using a binary mobile phase comprising solvent A (0.1% formic acid in UHP water at pH 2.73) and solvent B (acetonitrile). Satisfactory peak shapes and resolution were achieved using a gradient elution program consisting of mobile phase B (10%) for 0 - 1 min which was gradually (at 1% intervals) increased from 10% to 40

4. Method validation

Under optimised conditions, the developed method was validated using parameters such as linearity, limit of detection (LOD), limit of quantification (LOQ) and precision (intra-day and inter-day). Linearity was evaluated using calibration standards prepared by spiking ultrahigh purity (UHP) water at a concentration range of 0.5 - 2 000 µg L-1 and a satisfactory linearity was obtained from 2 - 1 000 µg L-1 with regression coefficients better than 0.9990. The LOD and LOQ values were calculated based on 3- and 10-times standard deviation of UHP water spiked with a minimum analyte concentration (0.3 - 0.5µg L-1) respectively. The LOD values ranged from 0.6 µg L-1 - 7.8 µg L-1 while LOQ values were obtained in the range of 1.8 - 23.4 µg L-1 (Table 1). 

Compound

Regression equation

(n = 6)

R2

Linear Range

(µg L-1)

LOD

(µg L-1)

LOQ

(µg L-1)

SGDy = 0.008x + 0.0450.99918 - 1 0002.78.1
SAMy =0.018x + 0.0170.99989 - 1 0006.419.2
SAAy = 0.057x + 0.1250.99993 - 1 0001.03.0
SDZy = 0.231x + 0.1980.99964 - 1 0002.26.6
STZy = 0.015x + 0.1100.99995 - 1 0001.85.4
SPYy = 0.054x + 0.1420.99993 - 1 0001.23.6
SMRy = 0.341x - 2.6310.99978 - 1 0007.823.4
SMTy = 0.133x + 0.2250.99982 - 1 0000.61.8
SMMy = 0.523x + 1.1180.99923 - 1 0000.92.7
SCPy= 0.273x - 0.2320.99913 - 1 0001.33.9
SMXy = 0.4034x - 0.7130.99982 - 1 0001.85.4
SSOy = 0.296x + 0.3880.99994 - 1 0001.75.1
SBZy = 0.310x + 0.3230.99965 - 1 0002.06.0
SQZy = 0.508x +   0.1010.99992 - 10000.61.8
SSAy = 0.051x +   0.4570.99994 - 10001.85.4

Table 1: Quantitative result of linearity, and LOD and LOQ values obtained for 15 SAs

Inter-day and intra-day precision was investigated by spiking UHP water samples with a mixture of target analytes at three concentration levels (40 µg L-1, 100 µg L-1, and 400 µg L-1) for three consecutive determinations in a single day and five determinations in five days, respectively. Intra-day %RSD values at concentration levels of 40 µg L-1, 100 µg L-1, and 400 µg L-1 were in the range 4.1 - 9.5%, 3.6 - 9.7%, and 1.5 - 9.3%, respectively, while for inter-day precision at the same three concentration levels, the ranges were 3.4 - 9.9%, 5.0 - 9.6%, and 3.1 - 9.9%, respectively (Table 2).

CompoundIntra-day (%RSD) (n = 6)Inter-day (%RSD) (n = 6)
40 (µg L-1)100 (µg L-1)400 (µg L-1)40 (µg L-1)100 (µg L-1)400 (µg L-1)
SGD9.78.18.99.19.65.9
SAM8.49.67.96.38.09.0
SAA9.29.78.35.08.09.9
SDZ8.78.83.28.07.56.4
STZ7.09.59.39.78.59.5
SPY9.09.47.85.86.89.2
SMR9.18.32.49.58.95.2
SMT9.28.96.93.48.33.8
SMM5.84.13.27.26.93.1
SCP9.57.77.98.68.87.0
SMX8.54.24.69.75.06.1
SSO9.06.97.99.68.07.9
SBZ9.38.83.29.06.69.0
SQZ5.73.61.59.99.86.0
SSA4.15.08.37.29.44.7

Table 2. Intra-day and inter-day precision results (%RSD) for water samples spiked at 40 µg L-1, 100 µgL-1, and 400 µg L-1 levels of 15 SAs

The accuracy of the method was evaluated by spiking the same levels of sulfonamide standard mixtures in three different types of water samples (river water, groundwater, and tap water) and the percentage recoveries were calculated. The results shown in Table 3 revealed that satisfactory recoveries, ranging from 70.5 to 103.9% with %RSD values ranging from 0.5 to 9.8%, were obtained, with the exception of SGD in river water and groundwater where recoveries in the range of 54.3 - 84.9% at the two higher concentration levels (100 µg L-1 and 400 µg L-1) were obtained. This lower recovery of SGD may be attributed to its very low KOW value (-1.07). 

Table 3: The recovery (%R) result obtained by comparing the ratio of concentration of each analyte found in spiked river water, groundwater samples. tap water I (Pretoria, South Africa) and tap water II (Florida, South Africa) at 40 µg L-1,100 µg L-1, and 400 µg L-1 levels of a mixture of SAs with the spiked concentration and results were multiplied by 100.
Furthermore, in order to study the effect of sample matrix on the extraction process, the relative recovery (RR), which is defined as the peak area ratio of the natural water sample to the UHP water sample spiked with analytes at the same concentration levels (40 µg L-1, 100 µg L-1, and 400 µg L-1) were calculated and found in the range of 70.3 - 106.1%, 72.5 - 105.4%, and 72.6 - 104.9%, respectively (Table S2, Supplemental material). The recovery results were in agreement with those previously reported in the literature on the same sample and matrix.34 The result demonstrated that the matrices of the real water samples do not have significant effects on the proposed DLLME method for the extraction of target analytes from real water samples. The selectivity of the proposed method was also evaluated by comparing the chromatograms of the blank (unspiked) water samples with the corresponding spiked water samples. Results shown in Figures S1 – S5 (supplemental material) confirmed that no interferences were observed at the retention times of the target analytes.
5. Application of the proposed method to Wastewater samples from Wastewater Treatment Plant (WWTP) 
To demonstrate the applicability of the proposed DLLME method, influent and effluent samples of the WWTP were collected, i.e., influent I, influent II, and effluent samples; these samples were analysed to assess the presence and levels of sulfonamides. A parallel group of the same samples was also analysed after spiking with 100 µg L-1 of target compounds. As shown in Table 4, SSO was detected in the influent I sample at an average concentration of 17.8 µg L-1 with corresponding %RSD, (n = 5) value of 1.6% and in the influent II sample at 14.8 µg L-1 with corresponding %RSD (n = 5) value of 1.9%. 

 

Compound

Influent IInfluent IIEffluent
CoCs%R%RSDCoCs%R%RSDCoCS%R%RSD
SGDND75.975.95.5ND70.970.95.2ND83.183.13.8
SAMND93.393.32.9ND96.896.85.6ND96.896.85.8
SAAND97.897.84.3ND97.897.83.2ND81.181.14.4
SDZND98.598.53.4ND73.973.90.7ND81.181.11.8
STZND96.096.03.7ND93.893.85.9ND98.798.71.3
SPYND94.194.16.3ND98.898.82.7ND92.392.31.9
SMRND83.383.31.1ND82.482.43.9ND80.780.73.4
SMTND91.291.21.9ND92.992.93.5ND94.294.24.1
SMMND90.190.10.7ND86.086.01.8ND82.282.22.0
SCPND98.998.91.3ND99.699.60.9ND101.7101.71.6
SMXND99.199.11.9ND99.199.10.2ND100.3100.32.7
SSO17.8105.487.61.614.8107.392.51.9ND93.093.04.8
SBZND102.2102.22.6ND99.999.91.5ND101.5101.52.9
SQZND80.580.52.0ND81.381.34.4ND107.7107.72.0
SSAND80.780.71.1ND70.570.54.3ND85.485.42.8

Table 4: Sulfonamide levels found in wastewater samples from wastewater treatment plant (WWTP), Pretoria, South Africa and spike recovery (%R, %RSD, n = 5 and two injections for each sample)
The chromatograms of SSO in influent I and II samples are shown in Figures S6 and S7, respectively (supplemental material). Sulfisoxazole was also detected in the influent of WWTPs in the USA in the range of 3.2 to 22.1 ng L-1,(Spongberg and Witter 2008)which was lower than that detected in the current work. It has also been detected in  WWTP effluents in  Canada (19 - 34 µg L-1),(Rodríguez-Cabo, Rodríguez et al. 2011) in the USA (3.2 to 11.9 ng L-1),(Spongberg and Witter 2008) and in Japan (0.13 -1.6 ng L-1)(Díaz-Cruz, García-Galán et al. 2008) One of the  possible reasons for the presence of sulfisoxazole is its wide application in human medicine on its own or together with erythromycin for urinary tract infections.(Chung 2008) In addition,  the compound is also excreted mainly in urine in an uncharged form. On the other hand, none of the target compounds were detected in the effluent samples. 
Sulfonamides have been  detected in WWTPs in Europe, America, Asia, Australia,(Zhang and Li 2011) and in Africa.(Faleye, Adegoke et al. 2018) Among these, sulfamethoxazole was the most  frequently detected, followed by sulfamethazine, sulfapyridine, and sulfadiazine.(Faleye, Adegoke et al. 2018Zhang, Bai et al. 2023). The highest reported concentration of sulfamethoxazole in WWTP influent was 5 597 ng L-1 and in WWTP effluent it was 6 000 ng L-1 in Europe, America, Asia and Australia,(Díaz-Cruz, García-Galán et al. 2008). United nation reported minimum and maximum concentrations for Africa were in the range of 0.1 - 29.0 μg L-1respectively(Faleye, Adegoke et al. 2018). Sulfamethoxazole has been detected wastewater effluents ranging between 0.15 and 10 μg L−1 in African wastewater effluents.(Ncube, Nuapia et al. 2021) Considering South  Africa, 
the average concentration of sulfamethoxazole in surface water was  reported as 7.3 μg L-1(Faleye, Adegoke et al. 2018) .  Due to the combination of trimethoprim and sulfamethoxazole used for the treatment of Pneumocystis pneumonia in people with HIV/AIDS, sulfamethoxazole is widely used and hence the most commonly detected sulfonamide in surface water systems in South Africa. Neube etal (Ncube, Nuapia et al. 2021) also reported the presence of sulfonamides (sulfamethiozole, sulfamethazine and sulfamethoxaxole) in their study for the presence of antibiotics in a South African stream.

6. Comparison of the proposed method with other reported DLLME methods for sulfonamide compounds

The proposed method was also compared with the other reported DLLME methods based on the number of analysed analytes, volume of solvent consumption, extraction time, linearity range, recovery values, and LOD values. The results shown in Table 5 confirmed that the proposed method has a wider linear range, short extraction time (less than 5 min), a lower consumption of organic solvents than the methods reported in the literature,(Wen, Li et al. 2011Li, Li et al. 2016) a larger number of analytes, and a wide range of KOW values (-1.07 - 3.8). However, the percentage recovery and LOD values were similar to those reported by Wen et al.(Wen, Li et al. 2011) and  Wang et al  (Wang, Li et al. 2022) and higher than the values reported by Xu et al.(Xu, Su et al. 2011) Therefore, the results indicated that the proposed method is feasible to be used as an alternative method for the determination of Sulfonamides in water samples.

MethodMatrixAnalyzed no SulfonamidesSolvent types and amount (extraction, disperser)Extraction time

Linearity

(µg L-1)

Recovery 

(%)

LOD 

(µg L-1)

Reference
DLLME - CE-DADLake, pond and tap water55 mL of chlorobenzene, 800 µL of DMSO40 - 45 min20 - 57053.6 - 94.00.02 - 0.57[32]
IL-based MADLLME-HPLC – FDWater, honey milk and plasma6100 µL of [C6MIM][PF6], 0.75 mL of MeOH15 min0.1 - 595.0 - 107.70.011 - 0.018 (river water)[33]
LDS-SD-DLLME - DMEEnvironmental water4

200 µL of octanol

750 µL of MeOH

600 µL of MeCN

17 min

1.0 - 500

to

10 - 500

NR0.22-1.92[34]
UA-DLLME-HPLC-DADEnvironmental water and seafood7

500 µL C2H2Cl4

900µL MeCN

More than 13min5 - 500080.0 – 116.0%0.7–7.8[35]
         
DLLME - UHPLC - DADMineral and runoff water

11   and

14 Qs

685 µL of CHCl3,

1 250 µL of MeCN

15 min5 - 100082.0 -115.0%0.8 - 32.1[36]
(UA-DLLME) - UPLC-MS/MSwater13-More than 10 min-80.3% - 101.80.6 ng/L and 2.4 ng/L[37]
DLLME - HPLC -DADTap water, river water and wastewater15

400 µL of CH2Cl2

600 µL of MeCN

5 min2 - 100070.8 -103.90.6 - 7.8Current work

Table 5: Comparison of the proposed DLLME method for water samples with other reported similar methods

7. Conclusions

In this study, dispersive liquid-liquid microextraction combined with high performance liquid chromatography-diode array detection has been evaluated and successfully developed for the simultaneous extraction and determination of  15 sulfonamides in real water samples. The effects of various parameters on the extraction efficiency were    evaluated, and optimum conditions were established. Under optimised conditions, linearity was found in the range of 2 - 1 000 μg L-1 with regression coefficients better than 0.9990. The limit of detection (LOD) and limit of quantification (LOQ) values were in the range of 0.6 - 7.8 μg L-1 and 1.8 - 23.4 µg L-1 respectively. Intra-day (n = 6) and inter-day (n = 6) precision expressed as %RSD were in the range of 1.5 - 9.7% and 3.1 - 9.9%, respectively. 
The accuracy   of the method was also evaluated in samples from three different sources of water (river water, groundwater, and tap water) and found to range from 70.5 to 103.9% with %RSD values ranging from 0.5 to 9.8% for 14 target analytes, with the exception of SGD, where recoveries in the range of 53.8 - 84.9% were obtained in river water and groundwater, with the corresponding %RSD values in the range of 0.5 - 9.5%.
Satisfactory relative recoveries (RR) in the range of 70.3 - 106.1% obtained provided evidence that the proposed DLLME   method was not affected by the matrix effects. Therefore, the proposed method was applied for the determination of sulfonamides in wastewater   treatment plant samples. Both SMX and SSO were detected; however, SMX was not quantifiable and SSO was in the range of 14.8 -17.8 µg L-1 in WWTP influents with %RSD values in the range of 1.6 - 1.9%. In comparison with other reported methods, the proposed DLLME method is adequate to be     used as an alternative method for the determination of sulfonamides in water samples.
Data availability
The raw data required to produce these findings can be obtained from authors with reasonable request
Conflict of interests
All the authors declare that they have no conflict of interest
Acknowledgments
The authors acknowledge the University of South Africa for the financial support and providing
the laboratory facilities for the student research during his PhD studies. Moreover, the authors also
Acknowledge the University of Debre Berhan (Debre Berhan, Ethiopia) for sponsoring the PhD study.

References

Clearly Auctoresonline and particularly Psychology and Mental Health Care Journal is dedicated to improving health care services for individuals and populations. The editorial boards' ability to efficiently recognize and share the global importance of health literacy with a variety of stakeholders. Auctoresonline publishing platform can be used to facilitate of optimal client-based services and should be added to health care professionals' repertoire of evidence-based health care resources.

img

Virginia E. Koenig

Journal of Clinical Cardiology and Cardiovascular Intervention The submission and review process was adequate. However I think that the publication total value should have been enlightened in early fases. Thank you for all.

img

Delcio G Silva Junior

Journal of Women Health Care and Issues By the present mail, I want to say thank to you and tour colleagues for facilitating my published article. Specially thank you for the peer review process, support from the editorial office. I appreciate positively the quality of your journal.

img

Ziemlé Clément Méda

Journal of Clinical Research and Reports I would be very delighted to submit my testimonial regarding the reviewer board and the editorial office. The reviewer board were accurate and helpful regarding any modifications for my manuscript. And the editorial office were very helpful and supportive in contacting and monitoring with any update and offering help. It was my pleasure to contribute with your promising Journal and I am looking forward for more collaboration.

img

Mina Sherif Soliman Georgy

We would like to thank the Journal of Thoracic Disease and Cardiothoracic Surgery because of the services they provided us for our articles. The peer-review process was done in a very excellent time manner, and the opinions of the reviewers helped us to improve our manuscript further. The editorial office had an outstanding correspondence with us and guided us in many ways. During a hard time of the pandemic that is affecting every one of us tremendously, the editorial office helped us make everything easier for publishing scientific work. Hope for a more scientific relationship with your Journal.

img

Layla Shojaie

The peer-review process which consisted high quality queries on the paper. I did answer six reviewers’ questions and comments before the paper was accepted. The support from the editorial office is excellent.

img

Sing-yung Wu

Journal of Neuroscience and Neurological Surgery. I had the experience of publishing a research article recently. The whole process was simple from submission to publication. The reviewers made specific and valuable recommendations and corrections that improved the quality of my publication. I strongly recommend this Journal.

img

Orlando Villarreal

Dr. Katarzyna Byczkowska My testimonial covering: "The peer review process is quick and effective. The support from the editorial office is very professional and friendly. Quality of the Clinical Cardiology and Cardiovascular Interventions is scientific and publishes ground-breaking research on cardiology that is useful for other professionals in the field.

img

Katarzyna Byczkowska

Thank you most sincerely, with regard to the support you have given in relation to the reviewing process and the processing of my article entitled "Large Cell Neuroendocrine Carcinoma of The Prostate Gland: A Review and Update" for publication in your esteemed Journal, Journal of Cancer Research and Cellular Therapeutics". The editorial team has been very supportive.

img

Anthony Kodzo-Grey Venyo

Testimony of Journal of Clinical Otorhinolaryngology: work with your Reviews has been a educational and constructive experience. The editorial office were very helpful and supportive. It was a pleasure to contribute to your Journal.

img

Pedro Marques Gomes

Dr. Bernard Terkimbi Utoo, I am happy to publish my scientific work in Journal of Women Health Care and Issues (JWHCI). The manuscript submission was seamless and peer review process was top notch. I was amazed that 4 reviewers worked on the manuscript which made it a highly technical, standard and excellent quality paper. I appreciate the format and consideration for the APC as well as the speed of publication. It is my pleasure to continue with this scientific relationship with the esteem JWHCI.

img

Bernard Terkimbi Utoo

This is an acknowledgment for peer reviewers, editorial board of Journal of Clinical Research and Reports. They show a lot of consideration for us as publishers for our research article “Evaluation of the different factors associated with side effects of COVID-19 vaccination on medical students, Mutah university, Al-Karak, Jordan”, in a very professional and easy way. This journal is one of outstanding medical journal.

img

Prof Sherif W Mansour

Dear Hao Jiang, to Journal of Nutrition and Food Processing We greatly appreciate the efficient, professional and rapid processing of our paper by your team. If there is anything else we should do, please do not hesitate to let us know. On behalf of my co-authors, we would like to express our great appreciation to editor and reviewers.

img

Hao Jiang

As an author who has recently published in the journal "Brain and Neurological Disorders". I am delighted to provide a testimonial on the peer review process, editorial office support, and the overall quality of the journal. The peer review process at Brain and Neurological Disorders is rigorous and meticulous, ensuring that only high-quality, evidence-based research is published. The reviewers are experts in their fields, and their comments and suggestions were constructive and helped improve the quality of my manuscript. The review process was timely and efficient, with clear communication from the editorial office at each stage. The support from the editorial office was exceptional throughout the entire process. The editorial staff was responsive, professional, and always willing to help. They provided valuable guidance on formatting, structure, and ethical considerations, making the submission process seamless. Moreover, they kept me informed about the status of my manuscript and provided timely updates, which made the process less stressful. The journal Brain and Neurological Disorders is of the highest quality, with a strong focus on publishing cutting-edge research in the field of neurology. The articles published in this journal are well-researched, rigorously peer-reviewed, and written by experts in the field. The journal maintains high standards, ensuring that readers are provided with the most up-to-date and reliable information on brain and neurological disorders. In conclusion, I had a wonderful experience publishing in Brain and Neurological Disorders. The peer review process was thorough, the editorial office provided exceptional support, and the journal's quality is second to none. I would highly recommend this journal to any researcher working in the field of neurology and brain disorders.

img

Dr Shiming Tang

Dear Agrippa Hilda, Journal of Neuroscience and Neurological Surgery, Editorial Coordinator, I trust this message finds you well. I want to extend my appreciation for considering my article for publication in your esteemed journal. I am pleased to provide a testimonial regarding the peer review process and the support received from your editorial office. The peer review process for my paper was carried out in a highly professional and thorough manner. The feedback and comments provided by the authors were constructive and very useful in improving the quality of the manuscript. This rigorous assessment process undoubtedly contributes to the high standards maintained by your journal.

img

Raed Mualem

International Journal of Clinical Case Reports and Reviews. I strongly recommend to consider submitting your work to this high-quality journal. The support and availability of the Editorial staff is outstanding and the review process was both efficient and rigorous.

img

Andreas Filippaios

Thank you very much for publishing my Research Article titled “Comparing Treatment Outcome Of Allergic Rhinitis Patients After Using Fluticasone Nasal Spray And Nasal Douching" in the Journal of Clinical Otorhinolaryngology. As Medical Professionals we are immensely benefited from study of various informative Articles and Papers published in this high quality Journal. I look forward to enriching my knowledge by regular study of the Journal and contribute my future work in the field of ENT through the Journal for use by the medical fraternity. The support from the Editorial office was excellent and very prompt. I also welcome the comments received from the readers of my Research Article.

img

Dr Suramya Dhamija

Dear Erica Kelsey, Editorial Coordinator of Cancer Research and Cellular Therapeutics Our team is very satisfied with the processing of our paper by your journal. That was fast, efficient, rigorous, but without unnecessary complications. We appreciated the very short time between the submission of the paper and its publication on line on your site.

img

Bruno Chauffert

I am very glad to say that the peer review process is very successful and fast and support from the Editorial Office. Therefore, I would like to continue our scientific relationship for a long time. And I especially thank you for your kindly attention towards my article. Have a good day!

img

Baheci Selen

"We recently published an article entitled “Influence of beta-Cyclodextrins upon the Degradation of Carbofuran Derivatives under Alkaline Conditions" in the Journal of “Pesticides and Biofertilizers” to show that the cyclodextrins protect the carbamates increasing their half-life time in the presence of basic conditions This will be very helpful to understand carbofuran behaviour in the analytical, agro-environmental and food areas. We greatly appreciated the interaction with the editor and the editorial team; we were particularly well accompanied during the course of the revision process, since all various steps towards publication were short and without delay".

img

Jesus Simal-Gandara

I would like to express my gratitude towards you process of article review and submission. I found this to be very fair and expedient. Your follow up has been excellent. I have many publications in national and international journal and your process has been one of the best so far. Keep up the great work.

img

Douglas Miyazaki

We are grateful for this opportunity to provide a glowing recommendation to the Journal of Psychiatry and Psychotherapy. We found that the editorial team were very supportive, helpful, kept us abreast of timelines and over all very professional in nature. The peer review process was rigorous, efficient and constructive that really enhanced our article submission. The experience with this journal remains one of our best ever and we look forward to providing future submissions in the near future.

img

Dr Griffith

I am very pleased to serve as EBM of the journal, I hope many years of my experience in stem cells can help the journal from one way or another. As we know, stem cells hold great potential for regenerative medicine, which are mostly used to promote the repair response of diseased, dysfunctional or injured tissue using stem cells or their derivatives. I think Stem Cell Research and Therapeutics International is a great platform to publish and share the understanding towards the biology and translational or clinical application of stem cells.

img

Dr Tong Ming Liu

I would like to give my testimony in the support I have got by the peer review process and to support the editorial office where they were of asset to support young author like me to be encouraged to publish their work in your respected journal and globalize and share knowledge across the globe. I really give my great gratitude to your journal and the peer review including the editorial office.

img

Husain Taha Radhi

I am delighted to publish our manuscript entitled "A Perspective on Cocaine Induced Stroke - Its Mechanisms and Management" in the Journal of Neuroscience and Neurological Surgery. The peer review process, support from the editorial office, and quality of the journal are excellent. The manuscripts published are of high quality and of excellent scientific value. I recommend this journal very much to colleagues.

img

S Munshi

Dr.Tania Muñoz, My experience as researcher and author of a review article in The Journal Clinical Cardiology and Interventions has been very enriching and stimulating. The editorial team is excellent, performs its work with absolute responsibility and delivery. They are proactive, dynamic and receptive to all proposals. Supporting at all times the vast universe of authors who choose them as an option for publication. The team of review specialists, members of the editorial board, are brilliant professionals, with remarkable performance in medical research and scientific methodology. Together they form a frontline team that consolidates the JCCI as a magnificent option for the publication and review of high-level medical articles and broad collective interest. I am honored to be able to share my review article and open to receive all your comments.

img

Tania Munoz

“The peer review process of JPMHC is quick and effective. Authors are benefited by good and professional reviewers with huge experience in the field of psychology and mental health. The support from the editorial office is very professional. People to contact to are friendly and happy to help and assist any query authors might have. Quality of the Journal is scientific and publishes ground-breaking research on mental health that is useful for other professionals in the field”.

img

George Varvatsoulias

Dear editorial department: On behalf of our team, I hereby certify the reliability and superiority of the International Journal of Clinical Case Reports and Reviews in the peer review process, editorial support, and journal quality. Firstly, the peer review process of the International Journal of Clinical Case Reports and Reviews is rigorous, fair, transparent, fast, and of high quality. The editorial department invites experts from relevant fields as anonymous reviewers to review all submitted manuscripts. These experts have rich academic backgrounds and experience, and can accurately evaluate the academic quality, originality, and suitability of manuscripts. The editorial department is committed to ensuring the rigor of the peer review process, while also making every effort to ensure a fast review cycle to meet the needs of authors and the academic community. Secondly, the editorial team of the International Journal of Clinical Case Reports and Reviews is composed of a group of senior scholars and professionals with rich experience and professional knowledge in related fields. The editorial department is committed to assisting authors in improving their manuscripts, ensuring their academic accuracy, clarity, and completeness. Editors actively collaborate with authors, providing useful suggestions and feedback to promote the improvement and development of the manuscript. We believe that the support of the editorial department is one of the key factors in ensuring the quality of the journal. Finally, the International Journal of Clinical Case Reports and Reviews is renowned for its high- quality articles and strict academic standards. The editorial department is committed to publishing innovative and academically valuable research results to promote the development and progress of related fields. The International Journal of Clinical Case Reports and Reviews is reasonably priced and ensures excellent service and quality ratio, allowing authors to obtain high-level academic publishing opportunities in an affordable manner. I hereby solemnly declare that the International Journal of Clinical Case Reports and Reviews has a high level of credibility and superiority in terms of peer review process, editorial support, reasonable fees, and journal quality. Sincerely, Rui Tao.

img

Rui Tao

Clinical Cardiology and Cardiovascular Interventions I testity the covering of the peer review process, support from the editorial office, and quality of the journal.

img

Khurram Arshad

Clinical Cardiology and Cardiovascular Interventions, we deeply appreciate the interest shown in our work and its publication. It has been a true pleasure to collaborate with you. The peer review process, as well as the support provided by the editorial office, have been exceptional, and the quality of the journal is very high, which was a determining factor in our decision to publish with you.

img

Gomez Barriga Maria Dolores

The peer reviewers process is quick and effective, the supports from editorial office is excellent, the quality of journal is high. I would like to collabroate with Internatioanl journal of Clinical Case Reports and Reviews journal clinically in the future time.

img

Lin Shaw Chin

Clinical Cardiology and Cardiovascular Interventions, I would like to express my sincerest gratitude for the trust placed in our team for the publication in your journal. It has been a true pleasure to collaborate with you on this project. I am pleased to inform you that both the peer review process and the attention from the editorial coordination have been excellent. Your team has worked with dedication and professionalism to ensure that your publication meets the highest standards of quality. We are confident that this collaboration will result in mutual success, and we are eager to see the fruits of this shared effort.

img

Maria Dolores Gomez Barriga

Dear Dr. Jessica Magne, Editorial Coordinator 0f Clinical Cardiology and Cardiovascular Interventions, I hope this message finds you well. I want to express my utmost gratitude for your excellent work and for the dedication and speed in the publication process of my article titled "Navigating Innovation: Qualitative Insights on Using Technology for Health Education in Acute Coronary Syndrome Patients." I am very satisfied with the peer review process, the support from the editorial office, and the quality of the journal. I hope we can maintain our scientific relationship in the long term.

img

Dr Maria Dolores Gomez Barriga

Dear Monica Gissare, - Editorial Coordinator of Nutrition and Food Processing. ¨My testimony with you is truly professional, with a positive response regarding the follow-up of the article and its review, you took into account my qualities and the importance of the topic¨.

img

Dr Maria Regina Penchyna Nieto

Dear Dr. Jessica Magne, Editorial Coordinator 0f Clinical Cardiology and Cardiovascular Interventions, The review process for the article “The Handling of Anti-aggregants and Anticoagulants in the Oncologic Heart Patient Submitted to Surgery” was extremely rigorous and detailed. From the initial submission to the final acceptance, the editorial team at the “Journal of Clinical Cardiology and Cardiovascular Interventions” demonstrated a high level of professionalism and dedication. The reviewers provided constructive and detailed feedback, which was essential for improving the quality of our work. Communication was always clear and efficient, ensuring that all our questions were promptly addressed. The quality of the “Journal of Clinical Cardiology and Cardiovascular Interventions” is undeniable. It is a peer-reviewed, open-access publication dedicated exclusively to disseminating high-quality research in the field of clinical cardiology and cardiovascular interventions. The journal's impact factor is currently under evaluation, and it is indexed in reputable databases, which further reinforces its credibility and relevance in the scientific field. I highly recommend this journal to researchers looking for a reputable platform to publish their studies.

img

Dr Marcelo Flavio Gomes Jardim Filho

Dear Editorial Coordinator of the Journal of Nutrition and Food Processing! "I would like to thank the Journal of Nutrition and Food Processing for including and publishing my article. The peer review process was very quick, movement and precise. The Editorial Board has done an extremely conscientious job with much help, valuable comments and advices. I find the journal very valuable from a professional point of view, thank you very much for allowing me to be part of it and I would like to participate in the future!”

img

Zsuzsanna Bene

Dealing with The Journal of Neurology and Neurological Surgery was very smooth and comprehensive. The office staff took time to address my needs and the response from editors and the office was prompt and fair. I certainly hope to publish with this journal again.Their professionalism is apparent and more than satisfactory. Susan Weiner

img

Dr Susan Weiner