Vitamin B 12 Assay – are we on Track?

Research Article | DOI: https://doi.org/10.31579/2690-8794/045

Vitamin B 12 Assay – are we on Track?

  • Rateesh Sareen 1*
  • G N Gupta 2

1* D.N.B Pathology, Department of Pathology & Transfusion Medicine, Santokba Durlabhji Hospital, Jaipur.
2 Head of department, Department of Pathology & Transfusion Medicine, Santokba Durlabhji Hospital, Jaipur.

*Corresponding Author: Rateesh Sareen, D.N.B Pathology, Department of Pathology & Transfusion Medicine, Santokba Durlabhji Hospital, Jaipur

Citation: Sareen R., G N Gupta, (2020) Vitamin B 12 Assay – are we on Track? J, Clinical Medical Reviews and Reports. 2(8); DOI: 10.31579/2690-8794/045

Copyright: © 2020, Rateesh Sareen, This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Received: 23 September 2020 | Accepted: 01 October 2020 | Published: 06 October 2020

Keywords: vitamin B12; reference interval (RI); laboratories

Abstract

Vitamin B12 assay is one of the most frequently ordered tests particularly as a part of regular medical checkups. The deficiency is rampant in vegetarian population. It is of immense importance that laboratories establish their own reference interval (RI) of analyte specially Vitamin B12 as  a diagnosis of Vitamin B12 deficiency based on RI of kit insert inadvertently leads to unnecessary treatment or work up. A blind reliance on RI of kit insert should be discouraged as they do not take into account population characteristics and do not truly reflect RI specific to the population under study.

Introduction

The water soluble Vitamin B12 is required for red blood cell formation, DNA synthesis and neurological function in human body exists in two active forms methycobalamin and 5-deoxyadenosylcobalamin [1, 2, and 3].

Dietary Vitamin B12 is bound to protein and therefore requires release by action of hydrochloric acid and protease enzyme in the stomach [3]. The synthetic form of Vitamin B12 is a free form that does not require metabolism in stomach. It directly combines with the intrinsic factor and gets absorbed within distal ileum by receptor mediated endocytosis [4].

Vitamin B12 is assessed by serum or plasma levels and values below 175-250 pg/ml for adults are indicative of Vitamin B12 deficiency [3].

The literature evidence suggests that serum Vitamin B12 levels might not precisely reflect intracellular concentration [5]. Even the surrogate markers like homocysteine levels (value>13micro mol/L) [6] and methyl malonic acid levels (value>0.4 micro mol/L) have poor specificity [4]. The laboratory diagnosis of Vitamin B12 status is based on mathematical model that combines multiple markers of Vitamin B12 status into a single diagnostic indicator a four variable analysis to calculate a combined indicator of vitamin B12 status [7].

Equation

This formula transforms all four markers to single variable dependent on age and is called Fedosov’s wellness score. It has been recently used in Non-anemic healthy Swiss citizens [8].

In India with large number of population having vegetarian diet and with limited resources, cost constraints it is not possible to apply Fedosov’s wellness score. Majority of patients are screened by complete blood count parameters including MCV.

The International Federation of Clinical Chemistry mandates every laboratory to have their own reference limits [9]. But unfortunately most of the laboratories in India follow reference intervals established in western population. The reference interval can be seriously factitious considering the variation in Indian and Western populations due to changes in dietary habits, life style and other factors. There is paucity of literature with regard to reference interval of Vitamin B12. The study aims to establish the reference interval (RI) for Vitamin B12 in our own population.

Material and methods

We measure Vitamin B12 levels by VITROS® Chemiluminesce Detection system (Micro Well Technology) Ortho Clinical Diagnostics U.S. The assay uses calibrators traceable to the certified National Institute of Standards and technology reference interval. The retrospective data was analyzed after ensuring that daily quality control was acceptable on each day of testing. The study included all patients who underwent Serum Vitamin B12 concentration over past one year. The exclusion criteria were:

  • All minors < 18 years of age
  • Hemoglobin <12.6g/dl (Both male & female)
  • Hemoglobin Males >16.5 g/ dl & Females >16.0 g/dl
  • Mean corpuscular volume (MCV)  < 83fl and >101fl
  • WBC- Less than 4000// mm3 and more than 10000/ / mm3
  • Platelet count less than 1, 40,000/ / mm3 and more than 4, 50,000/ / mm3

The complete blood count was analyzed on Sysmex XN1000 hematology 6-part auto analyzer.

The prime objective of the study was to evaluate RI in adult population in western India thriving predominantly on vegetarian diet. A total of 7416 patients underwent Vitamin B12 assay estimation over one year period in the year 2018. After applying the exclusion criteria 2510 patients with normal Mean corpuscular volume (MCV) were studied to determine the reference interval of Vitamin B12 in Indian population.

The analysis showed that Vitamin B12 and MCV do not show statistically significant correlation over entire 7416 patients. (r=-0.07361, p < 0.0001 significant). Statistical analysis was done using Medical (online free trial version) to check normality by Shapiro-Francia test and calculation of reference interval done.

The double sided reference interval study of serum Vitamin B12 for all adults is summarized in table-1. The lower limit of RI was 159 and upper limit 848.50. We used non parametric log transformation method as the data was not normal having kurtosis and skewness.

Limitation

  • History of intake of VitaminsB12 was not available
  • Study is limited to patient seeking medical advice at tertiary care hospital.
  • History of drugs, alcohol which may have profound effect on Vitamin B12 level was not available.

Strength

  • Large sample size- Drug supplements and other confounding factors are minimized due to larger sample size
  • Proper statistical analysis CLSI guidelines
  • Exclusion criteria 
  • Age old dictum of MCV as surrogate marker of Vitamin B12 deficiency is used as one of the exclusion criteria

Discussion

Reference interval (RI) plays a pivotal role in clinical laboratory test interpretation and patient care because 80% of the medical decisions are based on laboratory test results [10].  The original concept of RI was first enunciated by Grasbeck and Saris in 1969 [11]. RI are the soul of laboratory tests because isolated laboratory test values do not have any importance without age and sex specific reference range. Clinical and laboratory standards constitute approved guideline (CLSI) that elaborately described methodology for estimation of RI [12]. Reference interval (RI) as defined by Ceriotti, “is an interval that, when applied to the population serviced by the laboratory correctly includes most of the subjects with characteristics similar to reference group and excludes the others [13]. The majority of RI used today is with reference to population targeting 95% confidence limits. It implies that 5% of the results from healthy population will fall outside the reported RI. Therefore, no RI is absolutely right or wrong per se. Most of the laboratories do not establish their own reference range and through fait accompli use the manufacturer’s RI provided to them. 

Whenever a laboratory introduce a new test in their menu, the accreditation bodies make it mandatory to verify or establish performance characteristics. The new protocol C28-A3 allows laboratories to validate reference interval with small number of samples. A laboratory can now use 20 samples for validating its RI. In case the number of outliers is less than 2, it is statistically valid for the laboratory to adopt the RI as their own. In case the number of outliers is more than 2 that are present outside RI range then another set of 20 samples must be collected. 

The RI (normal range, reference range) to be calculated by using any of the three methods 1) Normal distributed values, 2) Non parametrical percentile method and 3) Robust method. When the sample size is small (less than 20) robust method can be applied. For bigger sample size, the choice of method depends on the nature of distribution of data. Data which has high kurtosis and skewness are best examined by percentile method.

The first and foremost step is to test for outlier based on tests by Reedetal [14] or Tukey [15]. The tests detected outliers which have to be inspected by the investigator so that they are excluded from calculation. The next step is the test for normal distribution. If p value is higher than 0.05, it may be assumed that the data follows normal distribution hence ‘accept’ normality. If p value is less than 0.05then the hypothesis of normal distributed data is rejected. If the data does not have normal distribution it requires logarithmic transformation (positively skewed data). The results of our study are enumerated in underlying tables.

Table-1: summary statistics

The prevalence of Vitamin B12 deficiency based on our data was 49.56% .Our calculated RI for adult came out to be 159- 843.5 pg/ml, whereas the kit insert RI is 239-931 pg/ml. It signifies that the Vitamin B12 levels in our population are far low as compared to those of kit insert. Furthermore as the lowest detection limit of the instrument Vitros 3600 was 159 pg/ml it is possible that it could be further go down in case the detection limit is lower. More studies are required to further elaborate on the lower detection limit. The upper normal reference limit was low in our study from 931pg/ml to 843.5 pg/ml.

It is evident that on following the kit insert reference range there is over diagnosis of Vitamin B12 deficiency. Even if we go by mayo clinic RI 180-914ng/l, the kit insert RI was much higher. Considering the limitation of our analyzer the lower limit of normal RI could be further below 159 ng/l.

Conclusion

The study established the rationale of using laboratory’s own reference interval by laboratories which are calculated on the basis of local and regional demographics.

References

Dear Editorial Team, Clinical Medical Reviews and Reports. My experience with the journal was highly positive. The peer-review process was rigorous, constructive, and completed in a timely manner. The reviewers provided valuable comments that helped improve the quality and clarity of our manuscript. The editorial office was professional, responsive, and supportive throughout all stages of the publication process. Communication was clear and efficient, and any questions were addressed promptly. Overall, I found the journal to maintain high scientific standards and an excellent publication workflow. I would be pleased to consider submitting future work to this journal. Best wishes from, Elena Popa.

img

Dr Elena Popa

It was my pleasure to submit my testimonial concerning the Reviewer Board of our Scientific Journal “Brain and Neurological Disorders”. The Reviewers focused on some modifications and their contribution was helpful. The ladies of our Editorial Office were also supported my efforts. It was my honor to have such a co-operation and I am looking forward for more collaboration.

img

Dr Nikolaos Andreas Chrysanthakopoulos

Dear Grace Pierce, Editorial Coordinator of Journal of Clinical Research and Reports, Thank you for the speedy and efficient peer review process. I appreciate the fact that your peer reviewers do not take months to respond like with some other journals. I would also like to thank the editorial office for responding quickly to my questions. It is an excellent journal. I plan to submit more manuscripts in the future. Best wishes from, Robert W. McGee

img

Robert W McGee

Dear Grace Pierce, Editorial Coordinator of Journal of Clinical Research and Reports, Working with you and your team on our recent publication in JCRR has been a truly wonderful and enjoyable experience. The responses were prompt, and the reviewers were patient, constructive, and highly professional. One reviewer in particular gave me the feeling that a professor was carefully reading and commenting on my coursework, which was deeply touching. The entire process was straightforward and hassle‑free, with no tedious online forms to complete. I highly recommend this journal. Best wishes from, DR Aibing Rao, Head of R&D

img

Aibing Rao

I Appreciate the Opportunity to Share my Experience with the Journal of Clinical Research and Reports. The peer review process was timely and constructive, and the feedback provided helped improve the quality of our manuscript. The editorial office was professional, responsive, and supportive throughout the process, ensuring smooth communication and efficient handling of the submission. Overall, it was a positive experience collaborating with your team.

img

Kashani Mehdi

Dear Mercy Grace, Editorial Coordinator of Obstetrics Gynecology and Reproductive Sciences, We would like to express our gratitude for your help at all stages of publishing and editing the article. The editors of the magazine answer all the necessary questions and help at every stage. We will definitely continue to cooperate and publish other works in the Obstetrics Gynecology and Reproductive Sciences! Best wishes from, Alla Konstantinovna Politova,

img

Alla Konstantinovna Politova