Chat with usResearch Article | DOI: https://doi.org/10.31579/2693-4779/179
IAE, 175-28, Goan-ro 51 beon-gil, Baegam-myeon, Cheoin-gu, Yongin-sí, Gyeonggi-do, Korea
*Corresponding Author: Dongchan Lee, IAE, 175-28, Goan-ro 51 beon-gil, Baegam-myeon, Cheoin-gu, Yongin-sí, Gyeonggi-do, Korea.
Citation: Dongchan Lee, (2024), Concept Design of a Smart Haptic Suit for a Biological Signal Recognition and Feedback to Human Body Movements, Clinical Research and Clinical Trials, 9(2); DOI:10.31579/2693-4779/179
Copyright: © 2024, Dongchan Lee. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Received: 12 January 2024 | Accepted: 17 January 2024 | Published: 26 January 2024
Keywords: smart clothing; ergonomic signal monitoring; motion artifact; auxetic pattern electrode; re-modularization
The ultimate purpose of monitoring bio-signals in the form of individually tailored clothing is to monitor an unconstrained bio-signals during the wearer's physiological activities and provide freedom of movement, thereby eliminating inconvenience in workspace of HMI (Human-Machine Interaction) or HRI (Human-Robot Interaction), and healthcare life. In the form of smart clothing for human collaboration industries or everyday life, there is a need to provide new sensitive and high accurate functions to the wearer while also faithfully performing the functions of existing clothing. In this respect, auxetic pattern-based sensors and actuators can be said to be one of the most promising candidates to be combined with smart wearable technology. There is an emerging need to research the concepts of these new sensors and actuators and present a paradigm for clothing research on innovative smart haptic suits. Accordingly, there is a need for a concept of smart haptic system suitable to meet the conditions of sensors and actuators with improved sensitivity, conductivity, and receptivity that can have characteristics such as microporosity, flexibility, and large surface area of nano and micro web structures. When measuring a bio-signal sensing and motion feedback using an electro-active structural pattern, the signal waveform can be confirmed to have a similar signal waveform to the measured signal through direct visual evaluation, and muscle signals for each body part can be measured using a commercialized sensor. The same signal can be confirmed. Through Pearson's correlation analysis using statistical analysis techniques, signal data between nano and micro web structure-based respiration sensors and existing respiration sensors showed significant correlation with each other, indicating respiration obtained from nano and micro web-based sensors. A theoretical model which predicts the actuation stroke of the system based on the material properties of the auxetic SMM (Shape Memory Material) and bias components as well as the geometry of the metamaterial system was developed.
Bio-monitoring smart clothing for human-robot interaction, human-machine interaction and health care is a new type of smart haptic clothing that combines the human body, with BT and IT functions. By attaching various devices that can check biological signals, signals such as the wearer's breathing or heart rate or body movements can be detected, allowing the wearer's bio-signals to be checked in a personalized way in daily life. Bio-monitoring smart clothing is a type of smart clothing that is essential for future daily life, where interest in human interaction industries or health is expected to increase. Bio-monitoring systems are evolving from measuring bio-signals using bio-medical devices such as thermometers or pulse monitors to measuring bio-signals through sensors integrated into clothing or bands. This phenomenon is changing from existing bio-medical devices to smart haptic suits that can be worn in daily life as interest in health increases and the desire to check one's health status at any time increases. The wearable system consists of four key elements. These include a 'sensing system' that detects physical and chemical changes in the human body, a 'monitoring system' that provides measured biometric information, an 'analysis system' that presents new health indicators based on the analysis of transmitted data, and a health status message to the wearer. It is a feedback system that notifies changes, etc. Wearable technology [1]. soft robotics [2]. human-robot interaction [3,4] and human-machine interaction [5]. have sparked a growing interest in embedded soft devices. Accordingly, various methods have been proposed to manufacture soft sensors that can be integrated into smart clothing systems [6,7]. In the future, for bio-monitoring smart clothing to become popular in everyday life, it is important that the core elements of the wearable system develop evenly to provide the maximum performance required by consumers, but the selective combination of the core elements for this is also very important [8-12].
The most important thing is to make the sensing system wearable, that is, to make the sensor wearable. Most have three functions for acquiring bio-signals. One is to measure the respiratory rate using an EMG sensor, the other is to measure the electrocardiogram using a medical ECG sensor, and the third is a system to measure the degree of movement through a motion measurement sensor such as an IMU. It consists of a memory card that stores data and a system that transmits, analyzes, and provides feedback. However, it is a block-based model made from a combination of structure and electronic devices, rather than being completely integrated. Information such as speed and exercise distance input through the heart rate and IMU sensor is displayed on the computer in real time, allowing you to know what level of intensity you are exercising. In addition, to check whether the electrical resistance changes depending on the degree of elongation of the structure, the change in electrical resistance can be measured with a multimeter, and then the change in electrical resistance of the structure according to the elbow angle can be measured in real time while wearing an arm sleeve. The electrical resistance values appear different even at the same angle when the structure is stretched (arm flexion operation) and contracted (arm extension operation). The reason is that when the arm is bent, the sensor attached to the elbow is stretched and receives a forced force in the opposite direction to its original shape, so the change in resistance occurs slowly, and when the arm is extended again, the structural sensor returns to its original shape. Changes in electrical resistance can be used at high speeds. If the principle of this motion measurement sensor is applied to other body parts besides the arms, it can be used to predict the body motion size of that part through clothing. When monitoring ECG and EMG signals, electrodes woven or woven from conductive metal thread are used as bioelectrodes. The method of sensing the microcurrent coming from the heart and transmitting it to the output device is the same as the electrode commonly used in medical institutions, but it is meaningful in that it is human-friendly and can be worn at all times [13-19].
In this paper, we analyzed key element technologies for the development of smart haptic suits by dividing them into key elements in the development of bio-monitoring smart clothing and e-textiles, the type of sensing system, electrocardiogram measurement, respiration measurement, and motion measurement. As mentioned in the introduction, bio-monitoring systems are evolving from non-wearable systems such as thermometers or pulse meters to wearable systems such as clothing or bands. This is an inevitable choice to meet human needs for high acceptability, high sensitivity, and convenience of wearing in the concept of a smart haptic suit that can analyze behavior and check bio-signals in daily life and provide muscle support anytime and anywhere. We intend to compare and examine the feasibility and feasibility of use as a smart textile by manufacturing conductive nano and micro webs as textile sensors using nanotechnology and metal coating techniques. Nano and micro web structures have characteristics such as porosity, flexibility, and large surface area, which can greatly contribute to improving conductivity. This is suitable to meet the conditions of the sensor, so based on this, it is expected to be suitable as a textile sensor. It is expected. Through this, we measure biological signals such as EMG, ECG and respiratory rate, verify the quality of the signal by comparing and analyzing it with signals obtained through existing non-textile sensors, and provide an integrated soft actuator function based on this.
2.1 Design Concept of Soft Sensor for Smart Haptic Suit
If the optimal conditions for producing conductive nano- and micro-web textiles with conductive polymers using sensors are discovered, it will be possible to develop textiles that minimize clothing wearers' resistance and naturally perform their unique functions within a smart clothing system. Flexible and soft sensors are currently the most studied component of smart clothing, but most studies simply focus on flexibility, and research on constructing smart clothing systems using nanomaterials is minimal. Accordingly, it can be said that research is needed to develop nano and micro web textiles that are flexible, lightweight, and highly conductive as sensors. The ultimate purpose of monitoring biological signals in the form of clothing is to monitor the wearer's physiological activity time. This provides new functions to the wearer in the form of smart clothing for HMI, HRI and everyday life while faithfully performing the functions of existing clothing. There is a need to do it. In this respect, nano or micro materials can be said to be one of the most promising candidates to be combined with wearable technology. So far, metal materials have been mainly used in research on the use of conductive materials as clothing materials, but there is an emerging need to research new materials and present an innovative paradigm for clothing research
The bio-signal measurement module receives bio-signals through three electrodes, and includes a hardware filter to block various noises, an amplifier circuit to amplify the bio-signals, and DSP (Digital Signal Processing) to filter software noise signals and convert the signals into data. It consists of a processor and a Bluetooth module for wireless data transmission. Accordingly, by using nano or micro technology and metal coating techniques, conductive nano and micro webs were manufactured as textile sensors and used as smart textiles. Nano and micro web structures have characteristics such as microporosity, flexibility, and large surface area, which can contribute to improving conductivity. Since this is suitable to meet the sensor conditions, it is expected to be suitable as a textile sensor based on this, and the following research contents can be included in the development of a smart haptic suit: (a) Fabrication of conductive nano and micro web samples based on metallic nanomaterials and textile sensorization: Fabrication of conductive nano and micro web structures/Excellent electrical properties and improved tensile strength/Improved signal transmission ability of nano and micro web-based sensors/Nano and micro web structures Checking the washing and abrasion resistance of web-based sensors. (b) Fabrication of conductive polymer-based conductive nano- and micro-web samples and textile sensorization: Fabrication of composite-based conductive nano- and micro-web structures using various conductive materials / Fabrication of conductive polymer-based conductive nano- and micro-web structures using various conductive materials. (c) Clothing evaluation and textile-based bio-signal monitoring system: Wearing evaluation is conducted on dummies and subjects, and the suitability is verified using statistical techniques by comparing signals with existing bio-signal measurement devices/nano and micro web-based biometrics. Development of signal monitoring smart clothing/To develop nano- and micro-web-based electrocardiogram electrodes Resistance characteristics such as line resistance and sheet resistance for bio-signal recognition/High-soluble gauge ratio to develop nano- and micro-web-based respiration sensors
2.2 Design of Auxetic Tactile sensor
Auxetics are defined as materials or structures with the elastic property of negative Poisson’s ratio (NPR): when the material is stretched in one direction, it expands in one or more transverse directions as well [20,21]. This differs from most materials/structures that typically show a contraction in the perpendicular direction to the stretched direction. Wearable electronic devices can monitor various strains, which is beneficial to human healthcare, medical rehabilitation, soft robotics, human-robot interactions, and human–machine interactions. Strain sensors have received more attention owing to their good conformality, integrability and wearability. To guarantee widespread application, wire strain sensors should possess high sensitivity, high stretchability, and excellent durability. The sensitivity, indicated by the gauge factor (GF), is evaluated by the following equation.
(1)
where , , R, C, ε and v denote the relative resistance change, the relative capacitance change, real-time resistance, real-time capacitance, tensile strain, and Poisson’s ratio, respectively. When an increasing force (∆F) applied onto one of the sensing units, the capacitance or resistance is measured for each increment. The sensitivity S of a capacitance or resistance or magnetic flux can be expressed as,
(2)
Where is the amount of the change in the sensed-magnetic field before and after strain applied, and is the applied force change? Among the above-mentioned indicators, stretchability and durability are especially important because they ensure large stress loading and stable output of sensing signals after repeated deformation. Although resistance change increases sharply under a large strain, irreversible structural damage appears, and repeated cycling induces a viscoelastic effect, which results in potential fracture and poor durability. To achieve controllable high sensitivity as well as good stretchability and durability, a novel design concept of auxetic materials/structures is integrated with a magnetic field sensing system in this work. The sensitivity of our strain sensors was tuned by tailoring the geometric dimensions of the auxetic structure. Upon mechanical loading, the sensors showed a structural deformation effect (i.e., longitudinal stretching accompanied by transverse structural deformation), which resulted in good stretchability and durability. As the strain increased, continuous stretching produced high sensitivity. The typical auxetic structures are listed in Table 1 [22]. Three architectures were explored as the fundamental pattern. The structural Poisson’s ratio(ν) is calculated using eq. (3):
(3)
where εx and εy represent the x-direction strain and y-direction strain.
As shown in Figure 1-(b), Eco-flex (1:1 mixture of casting and hardening materials) and PDMS (10:1 mixture of casting and hardening materials) were mixed at a 1:1 ratio for four auxetic sensors shown in Figure 1-(a). To prevent curing rejection, it was important to thoroughly mix each solution separately before combining them. Afterwards, NdFeB particles were added in a 1:1 weight ratio to the mixed silicone solution. In a vacuum chamber, bubbles were removed from the resulting mixture, which was injected into a mold created through 3D printing. To avoid the formation of a pseudo membrane that could reduce the efficacy of the sensor, the solution was carefully injected using a syringe instead of simply pouring and molding it. Next, the mold containing the solution was placed in an oven at 55 degrees Celsius for 3 hours. Once cured, the material was removed from the mold using tweezers. At a 10% tensile strain, the sum of factors for the re-entrant hexagon was 77uT, for the modified re-entrant hexagon it was 90uT, for the rotating triangular plates it was 99uT, and for the rotating square plates it reached 169uT, as shown in Figure 1-(C). Furthermore, to assess the Poisson's ratio of each sample, the width and length of the samples were measured both before and after the tensile test. The calculated Poisson's ratios were -0.1, -0.12, -0.66, and -0.75, respectively. An auxetic structure requires less force to obtain the same sensing value as a regular rectangular sensor of the same size. Additionally, when stretching the sensor over the same length, an auxetic structure demands less force. In Figure 2-(A), it can be observed that normal samples required approximately force of 1.0 N to induce a displacement of 100uT, whereas the auxetic samples required only about 0.2 N of force to achieve the same displacement. Furthermore, as shown in Figure 2-(B), when subjected to the same amount of tensile force, it was evident that the auxetic samples required less force compared to the normal samples. This is advantageous for applying sensors to small joints such as finger joints. We prepared a 33mm auxetic sensor and a regular rectangular sensor, both fixed at the length of 5mm of each sensor, in the same configuration. The calculated sensitivity using equation (2) shows (110 µT)/ (0.2 N) = 550 for the auxetic sensor, and (25 µT)/ (0.2 N) = 125 for the conventional sensor [22].
Auxetic Structure | Shape | Poisson’s ratios | Young’s modulus |
Re-entrant hexagon [20,21] | |||
Modified Re-entrant hexagons (with diamond-shaped cells) | |||
Rotating Triangular Plates (Equilateral) [23] | |||
Rotating Square Plate [24] |
Table 1: Characteristics of typical auxetic structure
※ Properties of unit cell material: Es: Young’s modulus, v: Poisson’s ratio, kh: stiffness constant, n: no. of cell strut of re-entrant hexagon, B=2lsinθ, H=h+2lcosθ
Material | Density (kg/m3) | Young’s Modulus (MPa) | Poisson’s Ratio |
TPUs | 1200 | 15.31 | 0.48 |
Table 2: Material properties
(a)
(b)
Figure 1: Experimental schematics: (A) Fabrication process of auxetic sensors, (B) Illustration of auxetic sensors, (C) Sum of factors for tensile ratio and Poisson's ratio is -0.1, -0.12, -0.66, -0.75 for the sensors of auxetic patterns (re-entrant hexagon, modified re-entrant hexagon, Rotating Triangular Plates and Rotating Square Plate)
(a) (b)
Figure 2: Operation of sensors with an auxetic matrix and without an auxetic matrix under different external tensile forces (a) Magnetic Field Variation with Force, (b) Tensile rate Due to Force.
3.1 Design Concept of Auxetic SMM
The importance of wearable devices continues to increase due to their ability to provide real-time body feedback, offer a variety of functions, and provide a comfortable experience for the wearer. To improve wearability, an auxetic SMM actuator can be developed by manufacturing a shape memory material such as shape memory alloy (SMA) or shape memory polymer (SMP)with an auxetic structure that is lightweight and changes its physical shape and stiffness depending on temperature. This wearable actuator has the characteristic of increasing wearability and convenience for the wearer's 3D curved human body, and by applying the auxetic SMM wearable actuator as a wearable haptic, virtual senses are provided through thermal and mechanical feedback. can be experienced, and through this, the auxetic SMM actuator can be used as a wearable haptic. The transition from the martensite (low temperature) phase to the austenite (high temperature) phase in auxetic SMM does not occur instantaneously in a specific temperature range but proceeds gradually over a temperature range. Figure 3 shows the relationship between displacement and temperature, where the austenite start temperature is shown as As, the austenite end temperature is shown as Af, and the martensite start and end temperatures are shown as Ms and Mf, respectively. In the temperature range indicated by , the alloy consists of a mixture of austenite and martensite. As can be seen from this, there is virtually no change in length below As, and as SMM is heated, no further change in length occurs above Af. Similarly, substantially no change in length occurs above Ms immediately after cooling, and no further change in length occurs below Mf; Typically, there is significant hysteresis in the length-temperature curve (Figure 3, 4). The resistance within the shaded region between Rmin and Rmax can be used as an analog for the auxetic SMM temperature, and the resistance-temperature curve exhibits significant hysteresis. Therefore, the transformation rate between the two phases can be inferred based only on the resistance value without directly measuring the temperature.
Figure 3: Relationship between auxetic SMM stroke and temperature
(a) Force-stroke control (b) Displacement-stoke control
Figure 4: Relationship of force & displacement with auxetic SMM stroke: ① non-deformation state, ② force and deformation state
Auxetic shape memory material actuators can be described as a control and sensing mechanism, generating control information for the actuator using a control scheme that utilizes resistive feedback and changes in actuator resistance over time as the electrical signals are applied to the actuator. These control and sensing mechanisms perform position control functions, measure the load on the actuator, detect a signal collision or mechanical failure and system errors in the actuator, and provide compensation for the actuator. In a preferred control mechanism, measurement of the discharge time of a capacitor connected in parallel with the actuator is used to measure the resistance of the actuator. The model predicts the operating output based on the balance of forces created by the interaction between the auxetic SMM and the bias component and is based on several assumptions: (a) The auxetic geometry has only a kinematic effect on the deformation of the system and does not contribute to the internal energy in terms of the work done on the actuator. This means that the influence of factors such as friction and the structural hinges and weight of the structure is negligible. (b) The stress-strain behavior of the auxetic SMM component in the martensitic phase can be expressed as a trilinear model with each line representing the twin martensitic state (M), the transition state (T), and the twin martensitic state (DM), respectively (Figure 5). The austenite state can be represented by a linear model. (c) Upon activation/heating, the auxetic SMM component completely converts from martensite to austenite phase. This is usually the case for auxetic SMM systems with small cross-sectional areas, such as wires and strips, which are activated through the thermal effect. (d) The model is static, so the deformations of both the auxetic SMM and the bias components are elastic, fully recoverable, and unaffected by the activation rate. When
activated using an electric current, the transformation from martensite to austenite phase occurs almost instantaneously
3.2 Numerical Descriptions of Auxetic SMM
For a prototype of auxetic SMM actuator, it is necessary to develop a numerical expression that can be used to predict the actuation output of the system as a function of the material properties of the individual components and the overall geometry of system. This actuator consists of three components: a SMM, a bias component, and an auxetic material geometry. Each plays a distinct role in ensuring the overall functioning of the complex system. Auxetic SMM components guide actuation of the system, bias components provide reversibility upon deactivation and reusability of the actuator, and auxetic material geometry controls the actuation stroke and force output of the system in axial and transverse directions. These three components are shown in Figure 5. To maximize the actuation stroke output, the auxetic SMM component is placed at position lAB and the bias component is placed at position lCD. This means that upon activation, the contraction of the auxetic SMM component is counteracted by the stretching of the bias component imposed through the kinematic constraints of the auxetic material deformation mode. Since lAB > lCD, this also leads to the total positive operating stroke of the actuator. To quantify the operating stroke of the system, it is necessary to equilibrate the force and displacement of the auxetic SMM and bias components along the lAB direction. Because the auxetic SMM component is aligned along lAB, the force-displacement plots obtained from uniaxial loading of this component can be used directly, as shown in Figure 5. However, this is not the case for the bias component oriented perpendicular to lAB. Therefore, the force and displacement exerted by the bias component in the direction lAB required to maintain the equilibrium of the system must first be found through the kinematic laws that define the deformation of the auxetic material system. [25-28].
The pre-strain of the auxetic SMM component can be driven by the elastic bias component, which in this case is assumed to be a flexure hinge or compliant systems that deforms linearly over high strains. This means that during actuator assembly, the auxetic SMM component must first be introduced within the system with a predetermined length lSMM, followed by the elastic bias component. For the bias component to pre-stress the SMM component, the initial length lbias in the remainder of the bias component must be less than the length lCD corresponding to the length lAB of the auxetic SMM component. This condition can be expressed as:
(4)
and to integrate the bias component within the actuator, we ensure that the pre-strain applied to the bias component imparts sufficient force to balance the SMM and induce pre-strain in the latter component. The force-displacement relationship of the auxetic SMM component in its martensitic phase is represented by a tri-linear model. The SMM has an initial length of lSMA at rest and the model (shown in figure 3) may be expressed in terms of force (FAB ) and total length (lAB ) along the line between the points A and B as:
(5)
(6)
(7)
where kM, kT and kDM are the stiffness constants for the twinned martensitic state, transition state and detwinned martensitic state in the uniaxial experiments of SMM wire, respectively, while FT , dT , FDM and dDM represent the force (F) and displacement (d) thresholds required to switch from the first to second state (eqs. (5)–(6)) and the second to third state (eqs. (6)–(7)). These three lines represent the force-displacement behavior of the martensitic SMM in the twinned, transition and detwinned state respectively and are valid within the limits defined in the equations.
This numerical modelling of the force-displacement of martensitic auxetic SMM is identical to the analytical method to model uni-axial bias actuator behavior. The equations can be expressed in terms of stiffness constants, transformation displacements and forces rather than Young's modulus, strain and stress for various geometric patterns. In the case of auxetic SMM, these factors may easily be derived directly from basic stress–strain plots in the experimental parts. As shown in Figure 5, the force-displacement behavior of the bias component counteracting the auxetic SMM is represented by a non-linear model along the line AB. This is due to the fact that although the stiffness of the bias component is constant, it is aligned along the line CD, orthogonal to the line AB. The austenitic force-displacement behavior in Figure 5 is represented by the following linear model:
(8)
(9)
(10)
where kA represents the stiffness constant of the SMA in its austenitic phase. lbias is the initial length of the elastic bias component and kbias is the associated stiffness constant. . .
Eq. (10) is non-linear model for the bias component along the line AB, and and for the martensitic and austenitic phase of actuator can be used to find the corresponding cold and hot equilibrium global length of actuator. The maximum permissible actuation for a given bias component can be given by,
(11)
(b) Deformation of unit auxetic cell under tensile loading
Figure 5: A qualitative plot showing the forces and displacements acting along lAB for the SMM in martensitic and austenitic state and the opposing bias component. The equilibrium points at the SMM hot and cold states are also indicated
Figure 6: Sample schematics of auxetic SMM actuator for arm muscular skeletal motion
(a) Flowchart for exercise prediction
(b) Flowchart for movement assessment
Figure 7: Flowchart for feedback control of smart haptic suit
In the user-centered smart haptic development technology, there is a need for continued development of integrated technology that can transform electronic devices into clothing. To advance from block-based to embedded and finally to fiber-based technology, new materials such as EAP, nanotechnology, and new technologies such as electrospinning in textile converting technology are the next generation. Active research must be conducted to develop materials. For the elements that make up smart clothing, such as sensors, power supplies, and output devices, to operate efficiently, interconnection technology that considers compatibility, durability, and usability must also be developed. Currently, there are differences in the level of development of clothing-embedded technology for each component of smart clothing. Therefore, standardized standards must be established to achieve effective interconnection between each component. Smart haptic suits are fashion items that must satisfy users in both functional and emotional aspects. Therefore, user-centered technology that reflects the potential needs of consumers requires the integration and interconnection technologies mentioned above. Smart haptic suit using these smart materials can be broadly divided into passive and active types, the latter of which consists of a sensor that can detect environmental changes and an actuator that can respond, and can detect physiological changes in the living body through electrical, magnetic, and A sensor that detects or monitors with optical signals is built or installed in the form of an array chip on a thin polymer sheet, paper, fiber material, or special material sheet. A bio-signal sensitive actuator that operates according to detected bio-signals is built into or installed on the same sheet. In some cases, the detected bio-signal can be communicated with an external digital device to drive an actuator mounted on a separate structure.
This research was supported by the Technology Innovation Program (20019115) funded by the Korea Planning & Evaluation Institute of Industrial Technology (KEIT) and the Ministry of Trade, Industry, & Energy (MOTIE, Korea).
Clearly Auctoresonline and particularly Psychology and Mental Health Care Journal is dedicated to improving health care services for individuals and populations. The editorial boards' ability to efficiently recognize and share the global importance of health literacy with a variety of stakeholders. Auctoresonline publishing platform can be used to facilitate of optimal client-based services and should be added to health care professionals' repertoire of evidence-based health care resources.
Journal of Clinical Cardiology and Cardiovascular Intervention The submission and review process was adequate. However I think that the publication total value should have been enlightened in early fases. Thank you for all.
Journal of Women Health Care and Issues By the present mail, I want to say thank to you and tour colleagues for facilitating my published article. Specially thank you for the peer review process, support from the editorial office. I appreciate positively the quality of your journal.
Journal of Clinical Research and Reports I would be very delighted to submit my testimonial regarding the reviewer board and the editorial office. The reviewer board were accurate and helpful regarding any modifications for my manuscript. And the editorial office were very helpful and supportive in contacting and monitoring with any update and offering help. It was my pleasure to contribute with your promising Journal and I am looking forward for more collaboration.
We would like to thank the Journal of Thoracic Disease and Cardiothoracic Surgery because of the services they provided us for our articles. The peer-review process was done in a very excellent time manner, and the opinions of the reviewers helped us to improve our manuscript further. The editorial office had an outstanding correspondence with us and guided us in many ways. During a hard time of the pandemic that is affecting every one of us tremendously, the editorial office helped us make everything easier for publishing scientific work. Hope for a more scientific relationship with your Journal.
The peer-review process which consisted high quality queries on the paper. I did answer six reviewers’ questions and comments before the paper was accepted. The support from the editorial office is excellent.
Journal of Neuroscience and Neurological Surgery. I had the experience of publishing a research article recently. The whole process was simple from submission to publication. The reviewers made specific and valuable recommendations and corrections that improved the quality of my publication. I strongly recommend this Journal.
Dr. Katarzyna Byczkowska My testimonial covering: "The peer review process is quick and effective. The support from the editorial office is very professional and friendly. Quality of the Clinical Cardiology and Cardiovascular Interventions is scientific and publishes ground-breaking research on cardiology that is useful for other professionals in the field.
Thank you most sincerely, with regard to the support you have given in relation to the reviewing process and the processing of my article entitled "Large Cell Neuroendocrine Carcinoma of The Prostate Gland: A Review and Update" for publication in your esteemed Journal, Journal of Cancer Research and Cellular Therapeutics". The editorial team has been very supportive.
Testimony of Journal of Clinical Otorhinolaryngology: work with your Reviews has been a educational and constructive experience. The editorial office were very helpful and supportive. It was a pleasure to contribute to your Journal.
Dr. Bernard Terkimbi Utoo, I am happy to publish my scientific work in Journal of Women Health Care and Issues (JWHCI). The manuscript submission was seamless and peer review process was top notch. I was amazed that 4 reviewers worked on the manuscript which made it a highly technical, standard and excellent quality paper. I appreciate the format and consideration for the APC as well as the speed of publication. It is my pleasure to continue with this scientific relationship with the esteem JWHCI.
This is an acknowledgment for peer reviewers, editorial board of Journal of Clinical Research and Reports. They show a lot of consideration for us as publishers for our research article “Evaluation of the different factors associated with side effects of COVID-19 vaccination on medical students, Mutah university, Al-Karak, Jordan”, in a very professional and easy way. This journal is one of outstanding medical journal.
Dear Hao Jiang, to Journal of Nutrition and Food Processing We greatly appreciate the efficient, professional and rapid processing of our paper by your team. If there is anything else we should do, please do not hesitate to let us know. On behalf of my co-authors, we would like to express our great appreciation to editor and reviewers.
As an author who has recently published in the journal "Brain and Neurological Disorders". I am delighted to provide a testimonial on the peer review process, editorial office support, and the overall quality of the journal. The peer review process at Brain and Neurological Disorders is rigorous and meticulous, ensuring that only high-quality, evidence-based research is published. The reviewers are experts in their fields, and their comments and suggestions were constructive and helped improve the quality of my manuscript. The review process was timely and efficient, with clear communication from the editorial office at each stage. The support from the editorial office was exceptional throughout the entire process. The editorial staff was responsive, professional, and always willing to help. They provided valuable guidance on formatting, structure, and ethical considerations, making the submission process seamless. Moreover, they kept me informed about the status of my manuscript and provided timely updates, which made the process less stressful. The journal Brain and Neurological Disorders is of the highest quality, with a strong focus on publishing cutting-edge research in the field of neurology. The articles published in this journal are well-researched, rigorously peer-reviewed, and written by experts in the field. The journal maintains high standards, ensuring that readers are provided with the most up-to-date and reliable information on brain and neurological disorders. In conclusion, I had a wonderful experience publishing in Brain and Neurological Disorders. The peer review process was thorough, the editorial office provided exceptional support, and the journal's quality is second to none. I would highly recommend this journal to any researcher working in the field of neurology and brain disorders.
Dear Agrippa Hilda, Journal of Neuroscience and Neurological Surgery, Editorial Coordinator, I trust this message finds you well. I want to extend my appreciation for considering my article for publication in your esteemed journal. I am pleased to provide a testimonial regarding the peer review process and the support received from your editorial office. The peer review process for my paper was carried out in a highly professional and thorough manner. The feedback and comments provided by the authors were constructive and very useful in improving the quality of the manuscript. This rigorous assessment process undoubtedly contributes to the high standards maintained by your journal.
International Journal of Clinical Case Reports and Reviews. I strongly recommend to consider submitting your work to this high-quality journal. The support and availability of the Editorial staff is outstanding and the review process was both efficient and rigorous.
Thank you very much for publishing my Research Article titled “Comparing Treatment Outcome Of Allergic Rhinitis Patients After Using Fluticasone Nasal Spray And Nasal Douching" in the Journal of Clinical Otorhinolaryngology. As Medical Professionals we are immensely benefited from study of various informative Articles and Papers published in this high quality Journal. I look forward to enriching my knowledge by regular study of the Journal and contribute my future work in the field of ENT through the Journal for use by the medical fraternity. The support from the Editorial office was excellent and very prompt. I also welcome the comments received from the readers of my Research Article.
Dear Erica Kelsey, Editorial Coordinator of Cancer Research and Cellular Therapeutics Our team is very satisfied with the processing of our paper by your journal. That was fast, efficient, rigorous, but without unnecessary complications. We appreciated the very short time between the submission of the paper and its publication on line on your site.
I am very glad to say that the peer review process is very successful and fast and support from the Editorial Office. Therefore, I would like to continue our scientific relationship for a long time. And I especially thank you for your kindly attention towards my article. Have a good day!
"We recently published an article entitled “Influence of beta-Cyclodextrins upon the Degradation of Carbofuran Derivatives under Alkaline Conditions" in the Journal of “Pesticides and Biofertilizers” to show that the cyclodextrins protect the carbamates increasing their half-life time in the presence of basic conditions This will be very helpful to understand carbofuran behaviour in the analytical, agro-environmental and food areas. We greatly appreciated the interaction with the editor and the editorial team; we were particularly well accompanied during the course of the revision process, since all various steps towards publication were short and without delay".
I would like to express my gratitude towards you process of article review and submission. I found this to be very fair and expedient. Your follow up has been excellent. I have many publications in national and international journal and your process has been one of the best so far. Keep up the great work.
We are grateful for this opportunity to provide a glowing recommendation to the Journal of Psychiatry and Psychotherapy. We found that the editorial team were very supportive, helpful, kept us abreast of timelines and over all very professional in nature. The peer review process was rigorous, efficient and constructive that really enhanced our article submission. The experience with this journal remains one of our best ever and we look forward to providing future submissions in the near future.
I am very pleased to serve as EBM of the journal, I hope many years of my experience in stem cells can help the journal from one way or another. As we know, stem cells hold great potential for regenerative medicine, which are mostly used to promote the repair response of diseased, dysfunctional or injured tissue using stem cells or their derivatives. I think Stem Cell Research and Therapeutics International is a great platform to publish and share the understanding towards the biology and translational or clinical application of stem cells.
I would like to give my testimony in the support I have got by the peer review process and to support the editorial office where they were of asset to support young author like me to be encouraged to publish their work in your respected journal and globalize and share knowledge across the globe. I really give my great gratitude to your journal and the peer review including the editorial office.
I am delighted to publish our manuscript entitled "A Perspective on Cocaine Induced Stroke - Its Mechanisms and Management" in the Journal of Neuroscience and Neurological Surgery. The peer review process, support from the editorial office, and quality of the journal are excellent. The manuscripts published are of high quality and of excellent scientific value. I recommend this journal very much to colleagues.
Dr.Tania Muñoz, My experience as researcher and author of a review article in The Journal Clinical Cardiology and Interventions has been very enriching and stimulating. The editorial team is excellent, performs its work with absolute responsibility and delivery. They are proactive, dynamic and receptive to all proposals. Supporting at all times the vast universe of authors who choose them as an option for publication. The team of review specialists, members of the editorial board, are brilliant professionals, with remarkable performance in medical research and scientific methodology. Together they form a frontline team that consolidates the JCCI as a magnificent option for the publication and review of high-level medical articles and broad collective interest. I am honored to be able to share my review article and open to receive all your comments.
“The peer review process of JPMHC is quick and effective. Authors are benefited by good and professional reviewers with huge experience in the field of psychology and mental health. The support from the editorial office is very professional. People to contact to are friendly and happy to help and assist any query authors might have. Quality of the Journal is scientific and publishes ground-breaking research on mental health that is useful for other professionals in the field”.
Dear editorial department: On behalf of our team, I hereby certify the reliability and superiority of the International Journal of Clinical Case Reports and Reviews in the peer review process, editorial support, and journal quality. Firstly, the peer review process of the International Journal of Clinical Case Reports and Reviews is rigorous, fair, transparent, fast, and of high quality. The editorial department invites experts from relevant fields as anonymous reviewers to review all submitted manuscripts. These experts have rich academic backgrounds and experience, and can accurately evaluate the academic quality, originality, and suitability of manuscripts. The editorial department is committed to ensuring the rigor of the peer review process, while also making every effort to ensure a fast review cycle to meet the needs of authors and the academic community. Secondly, the editorial team of the International Journal of Clinical Case Reports and Reviews is composed of a group of senior scholars and professionals with rich experience and professional knowledge in related fields. The editorial department is committed to assisting authors in improving their manuscripts, ensuring their academic accuracy, clarity, and completeness. Editors actively collaborate with authors, providing useful suggestions and feedback to promote the improvement and development of the manuscript. We believe that the support of the editorial department is one of the key factors in ensuring the quality of the journal. Finally, the International Journal of Clinical Case Reports and Reviews is renowned for its high- quality articles and strict academic standards. The editorial department is committed to publishing innovative and academically valuable research results to promote the development and progress of related fields. The International Journal of Clinical Case Reports and Reviews is reasonably priced and ensures excellent service and quality ratio, allowing authors to obtain high-level academic publishing opportunities in an affordable manner. I hereby solemnly declare that the International Journal of Clinical Case Reports and Reviews has a high level of credibility and superiority in terms of peer review process, editorial support, reasonable fees, and journal quality. Sincerely, Rui Tao.
Clinical Cardiology and Cardiovascular Interventions I testity the covering of the peer review process, support from the editorial office, and quality of the journal.
Clinical Cardiology and Cardiovascular Interventions, we deeply appreciate the interest shown in our work and its publication. It has been a true pleasure to collaborate with you. The peer review process, as well as the support provided by the editorial office, have been exceptional, and the quality of the journal is very high, which was a determining factor in our decision to publish with you.
The peer reviewers process is quick and effective, the supports from editorial office is excellent, the quality of journal is high. I would like to collabroate with Internatioanl journal of Clinical Case Reports and Reviews journal clinically in the future time.
Clinical Cardiology and Cardiovascular Interventions, I would like to express my sincerest gratitude for the trust placed in our team for the publication in your journal. It has been a true pleasure to collaborate with you on this project. I am pleased to inform you that both the peer review process and the attention from the editorial coordination have been excellent. Your team has worked with dedication and professionalism to ensure that your publication meets the highest standards of quality. We are confident that this collaboration will result in mutual success, and we are eager to see the fruits of this shared effort.
Dear Dr. Jessica Magne, Editorial Coordinator 0f Clinical Cardiology and Cardiovascular Interventions, I hope this message finds you well. I want to express my utmost gratitude for your excellent work and for the dedication and speed in the publication process of my article titled "Navigating Innovation: Qualitative Insights on Using Technology for Health Education in Acute Coronary Syndrome Patients." I am very satisfied with the peer review process, the support from the editorial office, and the quality of the journal. I hope we can maintain our scientific relationship in the long term.
Dear Monica Gissare, - Editorial Coordinator of Nutrition and Food Processing. ¨My testimony with you is truly professional, with a positive response regarding the follow-up of the article and its review, you took into account my qualities and the importance of the topic¨.
Dear Dr. Jessica Magne, Editorial Coordinator 0f Clinical Cardiology and Cardiovascular Interventions, The review process for the article “The Handling of Anti-aggregants and Anticoagulants in the Oncologic Heart Patient Submitted to Surgery” was extremely rigorous and detailed. From the initial submission to the final acceptance, the editorial team at the “Journal of Clinical Cardiology and Cardiovascular Interventions” demonstrated a high level of professionalism and dedication. The reviewers provided constructive and detailed feedback, which was essential for improving the quality of our work. Communication was always clear and efficient, ensuring that all our questions were promptly addressed. The quality of the “Journal of Clinical Cardiology and Cardiovascular Interventions” is undeniable. It is a peer-reviewed, open-access publication dedicated exclusively to disseminating high-quality research in the field of clinical cardiology and cardiovascular interventions. The journal's impact factor is currently under evaluation, and it is indexed in reputable databases, which further reinforces its credibility and relevance in the scientific field. I highly recommend this journal to researchers looking for a reputable platform to publish their studies.
Dear Editorial Coordinator of the Journal of Nutrition and Food Processing! "I would like to thank the Journal of Nutrition and Food Processing for including and publishing my article. The peer review process was very quick, movement and precise. The Editorial Board has done an extremely conscientious job with much help, valuable comments and advices. I find the journal very valuable from a professional point of view, thank you very much for allowing me to be part of it and I would like to participate in the future!”
Dealing with The Journal of Neurology and Neurological Surgery was very smooth and comprehensive. The office staff took time to address my needs and the response from editors and the office was prompt and fair. I certainly hope to publish with this journal again.Their professionalism is apparent and more than satisfactory. Susan Weiner