Can Body Height Be Used to Predict Knee Implant Sizes?

Research Article | DOI: https://doi.org/10.31579/2690-1919/040

Can Body Height Be Used to Predict Knee Implant Sizes?

  • Lisi Hu, Irrum Afzal, Roy Twyman, Deiary Kader (2020) Can Bo 1*
  • Roy Twyman 1
  • Deiary Kader 1

1 South West London Elective Orthopaedic Centre, London, United Kingdom.

*Corresponding Author: Irrum Afzal, South West London Elective Orthopaedic Centre, London, United Kingdom

Citation: Lisi Hu, Irrum Afzal, Roy Twyman, Deiary Kader (2020) Can Body Height Be Used to Predict Knee Implant Sizes? J Clinical Research and Reports, 3(1); DOI:10.31579/2690-1919/040

Copyright: © 2020 Irrum Afzal. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Received: 06 February 2020 | Accepted: 17 February 2020 | Published: 24 February 2020

Keywords: arthroplasty; replacement; knee; body height

Abstract

Introduction: Accuracy of pre-operative templating using radiographs or joint replacement programmes is variable. Implant modularity has increased the number of sizes available and trays used per procedure. Multiple intra-operative implant size trials lead to longer surgery time and greater instrument maintenance costs.

It was hypothesised that patient factors could predict the size of implant and tray required, thereby increasing efficiency and reducing operative time.

This study aims to identify correlation between implant size and body height to improve efficiency and reduce costs.

Method: This prospective cohort study includes patients who underwent primary total knee replacement using the DePuy ATTUNE® Primary Total Knee System between 1st January 2016 to 7th August 2017 performed by a single surgeon, at a single hospital. Post-operative x-rays were reviewed for appropriate implant sizing.

The DePuy ATTUNE® Primary Total Knee System has five sets, split into sizes 1-2, 3-5, 6-8 and 9-10.

Results: 188 patients (205 knees) were included, 66 male and 122 female. Male height was 174.6cm (152-194.3) with average implant size 8. Female height was 158.7cm (145-177.8) with average implant size 5. The Spearman rank correlation between body height, femoral and tibial size were 0.793 and 0.837 respectively. All men <170cm and >185cm used the 6-8 and 9-10 set respectively. All women <150cm and >170cm used the 3-5 and 6-8 set respectively.

Conclusion: The positive correlation identified coincides with existing literature. Using extremes of height and gender, surgeons and theatre staff could predict the likely trial set required, improve theatre efficiency and reduce costs.

Introduction

Pre-operatively planning in orthopaedic surgery can help anticipate problems that may arise intraoperatively and plan for appropriate equipment, therefore improving surgical efficiency and patient care (1). As part of planning for Total Knee Replacements (TKR), predicting the correct femoral and tibial component is an important step for surgeons. Accurate component size selection can aid proper knee biomechanics, therefore decreasing pain and the need for revision surgery, as well as the appropriate management of theatre stock, reducing intra-operative size trials, operating time and its associated complications (2). In 2014-2015, 2.3 million surgical instruments were sterilised in a single NHS trust, costing £2,484,000 per annum (3). Accurate component size prediction can reduce the number of size trials and surgical equipment trays being opened, therefore reducing sterilisation costs.

Traditional templating methods, using pre-operative plain radiographs and acetate, or digital templating using joint replacement computer programmes have variable accuracy (2). More recently, the use of three-dimensional templating or patient-specific instrumentation has shown very high accuracy for the prediction of actual implant size used, but is more costly in time and money (4). However, these templating methods should only be used as a guide for initial size trialling intra-operatively.

Implant modularity have increased the number of femoral and tibial component sizes and the number of trays used per procedure. Sizing is important in ligamentous balancing, allowing stability throughout full range of motion. Flexion space too small can lead to reduced range of motion or wear of the polyethylene insert, while a space too large may result in mid-flexion instability (5).

The DePuy ATTUNE® Primary Total Knee System has 10 femoral and tibial component sizes available, with an additional four narrow femoral component sizes, to meet the needs of the diverse worldwide population. There are five trays for the trial sizes. The femoral component trays are split into sizes 3-5 and 6-8, while the tibial components are grouped in one tray with sizes 3-8. There are two combination trays of both femoral and tibial components, in sizes 1-2 and 9-10 (6).

It has been thought that patient variables such as gender and body height can correlate to the implant size, and can be used as a part of pre-operatively planning.

The aim of this study was to identify if there was correlation between body height and the size of the ATTUNE® Primary Total Knee System, which may aid pre-operative planning and surgical efficiency.

Materials and Methods

This was a retrospective analysis of prospectively collected comprehensive data set. This study was undertaken at a single hospital, using single surgeon data. All patients who underwent a primary TKR using the ATTUNE® Primary Total Knee System under the care of this single surgeon from 1st January 2016 to 7th of August 2017 were identified and included in this study. Post-operative x-rays were reviewed for appropriate sizing of implants.

A total of 188 patients were included, 17 underwent bilateral knee replacements, resulting in a total of 205 knee replacements. The average age was 73.3 (53-89). There were no other exclusion criteria.

The implant sizes were recorded on the local hospital operating theatre software, Bluespier.

All the implant sizes recorded on Bluespier were validated against the femoral and tibial implant stickers recorded in the implant books. No discrepancies were identified.

The body height for all 205 knee replacements were obtained from the Pre-Operative Assessment, recorded on Bluespier, and recorded in centimetres (cms) to allow for consistency in the study.

R2 coefficients were calculated using the Spearman Rank Correlation between each of the femoral and tibial components to determine if there was a correlation between body height and implant size.

This study was an audit approved by the local trust Research and Development team and did not require ethical approval.

Results

188 patients (205 knees) were identified from the inclusion and exclusion criteria, 66 male and 122 female. Body height were recorded for each knee replacement, allowing for slight changes in body height for the same patient over time. The average male height was 174.6 (152-194.3) and average female height was 158.7 (145-177.8).

GenderFemoral size rangeFemoral size medianFemoral size modeTibial size rangeTibial size medianTibial size mode 
Male5-10876-1087 
Female3-8553-755 

Table 1. Femoral and tibial implant sizes split by gender

Figure 1. Correlation between body height and femoral component size

Figure 2. Correlation between body height and tibial component size

Table 2. R2 coefficients for tibial and femoral components when compared to gender based height

Table 3. Percentage of femoral implants used, split by sizes, in females

Table 4. Percentage of femoral implants used, split by sizes in males

Table 5. Estimated femoral tray required when separated by gender and height

Our study illustrated a positive correlation between body height and the ATTUNE® Primary Knee implant sizes. The strongest correlation was seen between body height and the tibial component.

When separated by gender, a positive correlation was also demonstrated in all groups. Men required larger components than women in the same height group, where on average men were one size bigger. Of the 205 knee replacements, no patients required the use of size 1-2 kit, and only men required components from the size 9-10 kit.

Discussion

Our data was limited to a single surgeon at a single hospital, producing a small cohort of patients. The patient population referred to this single hospital may demonstrate an uneven distribution of ethnicitiy and Body Mass Index compared to the general population, therefore our data may not be transferable to other centres. However, we took steps to ensure the accuracy of the data captured, including using two sources to identify the size of implants used.

This study identified one outlier, who was a male patient with a height of 171cm and required a size five femoral implant, the only male patient who required the size 3-5 femoral tray to be opened. Therefore it is worth noting that the correlation between height and implant size is a guide, and surgeons and theatre staff should be prepared to open kits which may not have been anticipated based on patient height and gender.

Hernandez-Vaquero et al., conducted a double-blind study evaluating the accuracy of templating 50 primary TKRs using the same surgeon and implant (Triathlon® Knee System (Stryker, Mahwah, New Jersey, United States)) with eight femoral and tibial sizes each. They found that using traditional radiographic methods alone, correct templating of the femoral and tibial components to the implanted occurred in 55% and 50% respectively, and sizing to within one size were 90% and 94% respectively (7).

Fawzy et al., correlated body height with the Oxford unicompartmental knee implant (four sizes available), concluding that height based on gender is a reliable method for predicting the femoral component size in 100 patients(8).

Trainer et al., compared shoe size and height with the DePuy PFC Sigma or DePuy Attune implant size in 100 primary TKRs by a single surgeon, and found shoe size to be a better predictor in determining implant size compared to height, particularly the tibial component (9).

Shershon et al., reviewed 3377 primary TKRs using implants from six different manufacturers, and found a linear correlation between height, weight and gender for implant sizes and performed a regression analysis, showing that using radiographic templating alongside demographic data allowed prediction within one size to 90%-99% for femoral and tibial components, and is superior to using demographic data alone (1).

While our findings confirm the existing literature regarding the correlation between implant size and patient demographics, we also looked at the data with a view to improve theatre efficiency and reduce costs.

Conclusion

We can use gender and the extremes of height range to help surgeons and theatre staff predict which size trial set is likely to be used, and therefore limit the number of unnecessary trays being opened, subsequently reducing theatre time, equipment maintenance and sterilization costs and improving theatre efficiency.

The relationship between implant size and body height can be used to facilitate equipment management and streamline intraoperative efficiency by using it to predict the operative kit and size of implant likely to be used, and therefore ensure it is in stock and on standby. Body height can be used alongside radiographic templating to ensure the most accurate method to determine implant size is used pre-operatively for best patient outcomes and surgical efficiency.

Acknowledgement

With thanks to Mohamed Tarek and Mohamed Alrashdan for their assistance in data collection.

Patient Consent

Patient consent was not required for the purposes of this study.

Disclosure statement

The authors report no conflict of interest concerning the materials or methods used in this study or the findings specified in this paper. The corresponding author is responsible for the accuracy and completeness of the submitted information.

References

Dear Editorial Team, Clinical Medical Reviews and Reports. My experience with the journal was highly positive. The peer-review process was rigorous, constructive, and completed in a timely manner. The reviewers provided valuable comments that helped improve the quality and clarity of our manuscript. The editorial office was professional, responsive, and supportive throughout all stages of the publication process. Communication was clear and efficient, and any questions were addressed promptly. Overall, I found the journal to maintain high scientific standards and an excellent publication workflow. I would be pleased to consider submitting future work to this journal. Best wishes from, Elena Popa.

img

Dr Elena Popa

It was my pleasure to submit my testimonial concerning the Reviewer Board of our Scientific Journal “Brain and Neurological Disorders”. The Reviewers focused on some modifications and their contribution was helpful. The ladies of our Editorial Office were also supported my efforts. It was my honor to have such a co-operation and I am looking forward for more collaboration.

img

Dr Nikolaos Andreas Chrysanthakopoulos

Dear Grace Pierce, Editorial Coordinator of Journal of Clinical Research and Reports, Thank you for the speedy and efficient peer review process. I appreciate the fact that your peer reviewers do not take months to respond like with some other journals. I would also like to thank the editorial office for responding quickly to my questions. It is an excellent journal. I plan to submit more manuscripts in the future. Best wishes from, Robert W. McGee

img

Robert W McGee

Dear Grace Pierce, Editorial Coordinator of Journal of Clinical Research and Reports, Working with you and your team on our recent publication in JCRR has been a truly wonderful and enjoyable experience. The responses were prompt, and the reviewers were patient, constructive, and highly professional. One reviewer in particular gave me the feeling that a professor was carefully reading and commenting on my coursework, which was deeply touching. The entire process was straightforward and hassle‑free, with no tedious online forms to complete. I highly recommend this journal. Best wishes from, DR Aibing Rao, Head of R&D

img

Aibing Rao

I Appreciate the Opportunity to Share my Experience with the Journal of Clinical Research and Reports. The peer review process was timely and constructive, and the feedback provided helped improve the quality of our manuscript. The editorial office was professional, responsive, and supportive throughout the process, ensuring smooth communication and efficient handling of the submission. Overall, it was a positive experience collaborating with your team.

img

Kashani Mehdi

Dear Mercy Grace, Editorial Coordinator of Obstetrics Gynecology and Reproductive Sciences, We would like to express our gratitude for your help at all stages of publishing and editing the article. The editors of the magazine answer all the necessary questions and help at every stage. We will definitely continue to cooperate and publish other works in the Obstetrics Gynecology and Reproductive Sciences! Best wishes from, Alla Konstantinovna Politova,

img

Alla Konstantinovna Politova